Assessing wear of the acetabular cup using computed tomography: an ex vivo study.

作者信息

Olivecrona L, Jedenmalm A, Aspelin P, Weidenhielm L, Noz M E, Maguire G Q, Zeleznik M P, Olivecrona H

机构信息

Department of Radiology, Karolinska University Hospital Solna, Stockholm, Sweden.

出版信息

Acta Radiol. 2005 Dec;46(8):852-7. doi: 10.1080/02841850500271074.

Abstract

PURPOSE

To validate a clinically useful method for measuring acetabular cup wear using computed tomography (CT).

MATERIAL AND METHODS

Eight uncemented acetabular cups were scanned twice ex vivo using CT. The linear penetration depth of the femoral component head into the cup and the thickness of the remaining polyethylene liner were measured in the CT volumes using dedicated software. Two independent examiners twice assessed each volume. The CT measurements were compared to direct measurements using a coordinate measuring device and micrometer measurements.

RESULTS

Accuracy of wear measurements expressed as penetration depth was +/-0.6 and +/- 1.0 mm for the two examiners, respectively, with no significant differences between examiners, trials, and CT scans. Accuracy of measurements of remaining polyethylene was +/- 1.3 and +/- 1.0 mm, respectively, for the two examiners. Systematic differences between examiners were found, but no significant differences between trials and CT scans. These differences were due to different interpretations of metal artifacts in the volumes.

CONCLUSION

The proposed CT method for evaluating wear as head penetration depth allows for reliable wear detection at a clinically relevant level. Measurements of remaining polyethylene on CT volumes are not as reliable as wear measurements owing to metal artifacts.

摘要

目的

验证一种使用计算机断层扫描(CT)测量髋臼杯磨损的临床实用方法。

材料与方法

对八个非骨水泥型髋臼杯进行体外CT扫描,扫描两次。使用专用软件在CT容积中测量股骨部件头部进入髋臼杯的线性穿透深度以及剩余聚乙烯衬垫的厚度。两名独立的检查人员对每个容积进行两次评估。将CT测量结果与使用坐标测量装置的直接测量结果和千分尺测量结果进行比较。

结果

以穿透深度表示的磨损测量精度,两名检查人员分别为±0.6毫米和±1.0毫米,检查人员、测量次数和CT扫描之间无显著差异。两名检查人员对剩余聚乙烯的测量精度分别为±1.3毫米和±1.0毫米。发现检查人员之间存在系统差异,但测量次数和CT扫描之间无显著差异。这些差异是由于对容积中金属伪影的不同解读所致。

结论

所提出的将磨损评估为头部穿透深度的CT方法能够在临床相关水平上实现可靠的磨损检测。由于金属伪影,在CT容积上对剩余聚乙烯的测量不如磨损测量可靠。