Suppr超能文献

选择离子流管质谱法在三丙酮三过氧化物实时检测中的应用。

Application of selected-ion flow tube mass spectrometry to the real-time detection of triacetone triperoxide.

作者信息

Wilson Paul F, Prince Barry J, McEwan Murray J

机构信息

Department of Chemistry, University of Canterbury, PB 4800, Christchurch, New Zealand.

出版信息

Anal Chem. 2006 Jan 15;78(2):575-9. doi: 10.1021/ac051501b.

Abstract

A fast, efficient, real-time method for the quantitative analysis of the peroxide explosive, TATP, is described. The method utilizes rapid ion-molecule reactions of chemical reagent ions with the vapor above solid TATP. The reactions of three reagent ions (H3O+, O2+, NO+) were examined. Although all three ions exhibited a near-collision-rate reaction with TATP, only NO+ showed product ions that provide unequivocal evidence for a TATP-based explosive. The limit of detection of TATP in the gas phase is 10 ppb (4 x 10(-10) mol L(-1)).

摘要

本文描述了一种用于定量分析过氧化物炸药三过氧化三丙酮(TATP)的快速、高效、实时方法。该方法利用化学试剂离子与固体TATP上方蒸汽的快速离子-分子反应。研究了三种试剂离子(H3O+、O2+、NO+)的反应。尽管所有三种离子与TATP都表现出近碰撞速率反应,但只有NO+显示出能为基于TATP的炸药提供明确证据的产物离子。气相中TATP的检测限为10 ppb(4×10(-10) mol L(-1))。