Atkinson Wendy, Lockhart Stephen J, Houghton Lesley A, Keevil Brian G
Department of Medicine, South Manchester University Hospitals NHS Trust, UK.
J Chromatogr B Analyt Technol Biomed Life Sci. 2006 Feb 17;832(1):173-6. doi: 10.1016/j.jchromb.2005.12.038. Epub 2006 Jan 24.
A sensitive and rapid assay is described for the measurement of low concentrations of 5-hydroxytryptamine (5-HT) present in human platelet-depleted plasma (PDP) using reverse-phase high performance liquid chromatography (HPLC) with fluorimetric detection. With an analysis time of 12 min, this method is particularly useful for large-scale clinical trials investigating small differences in PDP 5-HT concentrations in conditions such as functional gastrointestinal disorders (FGID). The limit of detection and quantification were 1 and 3 nmol/l, respectively, and the calibration curve linear between 1 and 1000 nmol/l. The within-day and between-day precision were 4.3 and <13.6%, respectively.
描述了一种灵敏且快速的测定方法,用于使用反相高效液相色谱(HPLC)和荧光检测法测量人去血小板血浆(PDP)中存在的低浓度5-羟色胺(5-HT)。该方法分析时间为12分钟,对于在诸如功能性胃肠疾病(FGID)等情况下研究PDP中5-HT浓度微小差异的大规模临床试验特别有用。检测限和定量限分别为1和3 nmol/l,校准曲线在1至1000 nmol/l之间呈线性。日内和日间精密度分别为4.3%和<13.6%。