Short-term microvascular response of striated muscle to cp-Ti, Ti-6Al-4V, and Ti-6Al-7Nb.

作者信息

Pennekamp Peter H, Gessmann Jan, Diedrich Oliver, Burian Björn, Wimmer Markus A, Frauchiger Vinzenz M, Kraft Clayton N

机构信息

Department of Orthopedic Surgery, University of Bonn, Sigmund-Freud-Strasse 25, D-53105 Bonn, Germany.

出版信息

J Orthop Res. 2006 Mar;24(3):531-40. doi: 10.1002/jor.20066.

Abstract

Due to excellent mechanical properties and good corrosion resistance, titanium-aluminium-vanadium (Ti-6Al-4V) and titanium-aluminium-niobium (Ti-6Al-7Nb) are extensively used for orthopedic surgery. Concern has been voiced concerning the implications of the constituent vanadium in Ti-6Al-4V on the surrounding environment. Particularly in osteosynthesis where the alloys stand in direct contact to skeletal muscle, undesirable biologic reactions may have severe consequences. In a comparative study, we assessed in vivo nutritive perfusion and leukocytic response of striated muscle to the metals Ti-6Al-4V, Ti-6Al-7Nb, and commercially pure titanium (cpTi), thereby drawing conclusions on their short-term inflammatory potential. In 28 hamsters, utilizing the dorsal skinfold chamber preparation and intravital microscopy, we quantified primary and secondary leukocyte-endothelial cell interaction, leukocyte extravasation, microvascular diameter change, and capillary perfusion in collecting and postcapillary venules of skeletal muscle. A manifest discrepancy between the metals concerning impact on local microvascular parameters was not found. All metals induced an only transient and moderate inflammatory response. Only a slight increase in leukocyte recruitment and a more sluggish recuperation of inflammatory parameters in animals treated with Ti-6Al-4V compared to the other two metals suggested a minor, overall not significant discrepancy in biocompatibility. Gross toxicity of bulk Ti-6Al-4V on surrounding tissue could not be found. Conclusively, the commonly used biomaterials Ti-6Al-4V, Ti-6Al-7Nb, and cpTi induce an only transient inflammatory answer of the skeletal muscle microvascular system. Our results indicate that on the microvascular level the tested bulk Ti-alloys and cpTi do not cause adverse biologic reactions in striated muscle.

摘要

由于具有出色的机械性能和良好的耐腐蚀性,钛铝钒(Ti-6Al-4V)和钛铝铌(Ti-6Al-7Nb)被广泛应用于骨外科手术。人们对Ti-6Al-4V中的钒成分对周围环境的影响表示担忧。特别是在骨接合术中,合金与骨骼肌直接接触,不良的生物反应可能会产生严重后果。在一项对比研究中,我们评估了横纹肌对金属Ti-6Al-4V、Ti-6Al-7Nb和商业纯钛(cpTi)的体内营养灌注和白细胞反应,从而得出它们短期炎症潜力的结论。在28只仓鼠中,利用背部皮褶腔制备和活体显微镜技术,我们对骨骼肌集合静脉和毛细血管后微静脉中的一级和二级白细胞-内皮细胞相互作用、白细胞渗出、微血管直径变化和毛细血管灌注进行了量化。未发现这些金属在对局部微血管参数的影响方面存在明显差异。所有金属仅引起短暂和中度的炎症反应。与其他两种金属相比,用Ti-6Al-4V处理的动物中白细胞募集仅略有增加,炎症参数的恢复也较为缓慢,这表明在生物相容性方面存在微小但总体不显著的差异。未发现块状Ti-6Al-4V对周围组织有明显毒性。总之,常用的生物材料Ti-6Al-4V、Ti-6Al-7Nb和cpTi仅引起骨骼肌微血管系统短暂的炎症反应。我们的结果表明,在微血管水平上,测试的块状钛合金和cpTi不会在横纹肌中引起不良生物反应。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验