Marmulla Rüdiger, Mühling Joachim
Department of Cranio-Maxillofacial Surgery, University of Heidelberg, Heidelberg, Germany.
J Oral Maxillofac Surg. 2006 Mar;64(3):466-70. doi: 10.1016/j.joms.2005.11.019.
Motion artifacts can significantly deteriorate the precision of a computer-assisted surgical intervention because they destroy the isometric representation of tomographic pictures. In the context of a study, the influence of typical motion artifacts on the precision of markerless laser registration in image-guided oral and maxillofacial surgery was analyzed, and quality factors for evaluation of the isometry of a computed tomography (CT) dataset were determined.
Twenty patients underwent markerless registration, the precision being determined by means of intraoral evaluation markers. Then the 20 CT datasets were used for simulation of a typical motion artifact. The precision of the overlay of the dataset was checked again on the navigation workstation, in absence of the patient. The navigation system used was the Surgical Segment Navigator SSN++ (University of Heidelberg, Heidelberg, Germany).
The motion artifacts reduced the average patient registration from 1 to 4 mm. Quality factors for the isometry of a CT dataset were: the volume enclosed between the soft tissue mantles of the preoperative CT dataset and the intraoperative laser scan dataset, as well as the orientation of the normal vectors on the 3-dimensional reconstruction of the CT dataset.
The isometry of a CT dataset should always be checked before performance of a computer assisted surgical intervention because anisometric datasets result in inaccurate patient registration and navigation.
运动伪影会严重降低计算机辅助手术干预的精度,因为它们会破坏断层图像的等距表示。在一项研究中,分析了典型运动伪影对图像引导口腔颌面外科中无标记激光配准精度的影响,并确定了用于评估计算机断层扫描(CT)数据集等距性的质量因素。
20例患者接受了无标记配准,精度通过口内评估标记来确定。然后使用这20个CT数据集模拟典型运动伪影。在没有患者的情况下,在导航工作站上再次检查数据集叠加的精度。使用的导航系统是手术段导航器SSN++(德国海德堡大学,海德堡)。
运动伪影使平均患者配准从1毫米降低到4毫米。CT数据集等距性的质量因素为:术前CT数据集和术中激光扫描数据集的软组织覆盖层之间的封闭体积,以及CT数据集三维重建上法向量的方向。
在进行计算机辅助手术干预之前,应始终检查CT数据集的等距性,因为不等距数据集会导致患者配准和导航不准确。