Mammo Layla, Sheereen Atia, Saour Tanya, Saour Jalal
Department of Biological and Medical Research, King Faisal Specialist Hospital and Research Center, PO Box 3354, Riyadh 11211, Saudi Arabia.
Clin Biochem. 2006 Jul;39(7):767-9. doi: 10.1016/j.clinbiochem.2006.01.019. Epub 2006 Mar 23.
To test reproducibility, speed and cost of testing for factor V Leiden and FII 20210G>A in our practice.
We compared conformity, reproducibility, speed and cost using the Light Cycler (LC) and PCR-RFLP.
There was 100% conformity and reproducibility. LC was faster but 23% more expensive per sample. When equipment depreciation and patient expenses are added, LC testing becomes cheaper.
In our practice, LC provides fast, reproducible and cost-effective results.