Suppr超能文献

使用256排探测器CT扫描仪的三种加权费尔德坎普算法的噪声特性:肝脏容积电影成像的案例研究

Noise properties for three weighted Feldkamp algorithms using a 256-detecotor row CT-scanner: case study for hepatic volumetric cine imaging.

作者信息

Mori Shinichiro, Endo Masahiro, Obata Takayuki, Kishimoto Riwa, Kato Hirotoshi, Kandatsu Susumu, Tsujii Hirohiko, Tanada Shuji

机构信息

Department of Medical Physics, National Institute of Radiological Sciences, Chiba 263-8555, Japan.

出版信息

Eur J Radiol. 2006 Aug;59(2):289-94. doi: 10.1016/j.ejrad.2006.02.007. Epub 2006 Mar 29.

Abstract

In cone-beam geometry, image quality may be degraded or artifacts may occur if the cone angle is substantially wide. This is because a cone-beam scan along a circular orbit does not collect the complete set of data required to make an exact reconstruction of all volumetric data. To increase temporal resolution and thus image quality in cone-beam geometry, Silver proposed the new half-scan algorithm (NHS-FDK), which extends Parker's weighting function (HS-FDK) by utilizing a larger range up to 2pi. Here, we evaluated these algorithms for hepatic contrast-enhanced CT in cine scan mode using a 256-detector row CT. The full-scan (FS-FDK) images show uniform distribution of the image noise and CT-number uniformity. Image noise and CT-number uniformity with HS-FDK and NHS-FDK images follow the initial projection angle. HS-FDK images therefore have more changeable higher intensity (brighter) and a lower intensity (darker) areas than respective FS-FDK and NHS-FDK images. We concluded that, considering the trade-off between image quality and temporal resolution, the NHS-FDK algorithm is useful in volumetric cine imaging for the abdomen.

摘要

在锥形束几何结构中,如果锥角过大,图像质量可能会下降或出现伪影。这是因为沿圆形轨道进行的锥形束扫描无法收集到对所有体数据进行精确重建所需的完整数据集。为了提高锥形束几何结构中的时间分辨率从而提高图像质量,西尔弗提出了新的半扫描算法(NHS-FDK),该算法通过利用更大的范围直至2π扩展了帕克的加权函数(HS-FDK)。在此,我们使用256排探测器CT在电影扫描模式下对肝脏对比增强CT的这些算法进行了评估。全扫描(FS-FDK)图像显示图像噪声分布均匀且CT值均匀。HS-FDK和NHS-FDK图像的图像噪声和CT值均匀性随初始投影角度而变化。因此,与各自的FS-FDK和NHS-FDK图像相比,HS-FDK图像具有更多变化的高强度(更亮)和低强度(更暗)区域。我们得出结论,考虑到图像质量和时间分辨率之间的权衡,NHS-FDK算法在腹部容积电影成像中是有用的。

相似文献

7
Respiratory correlated segment reconstruction algorithm towards four-dimensional radiation therapy using carbon ion beams.
Radiother Oncol. 2006 Sep;80(3):341-8. doi: 10.1016/j.radonc.2006.07.035. Epub 2006 Sep 1.
8
[Cone-beam pseudo lambda tomography under FDK framework].
Nan Fang Yi Ke Da Xue Xue Bao. 2009 Oct;29(10):2094-7.
9
A combination-weighted Feldkamp-based reconstruction algorithm for cone-beam CT.
Phys Med Biol. 2006 Aug 21;51(16):3953-65. doi: 10.1088/0031-9155/51/16/005. Epub 2006 Aug 2.

引用本文的文献

1
Multibeam field emission x-ray system with half-scan reconstruction algorithm.
Med Phys. 2010 Jul;37(7):3773-81. doi: 10.1118/1.3446801.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验