Miró R, Soler J, Gallardo S, Campayo J M, Díez S, Verdú G
Departamento de Ingeniería Química y Nuclear, Universidad Politécnica de València, Camí de Vera, s/n, 46022 València, Spain.
Radiat Prot Dosimetry. 2005;116(1-4 Pt 2):65-8. doi: 10.1093/rpd/nci125.
A Theratron 780 (MDS Nordion) 60Co radiotherapy unit has been simulated with the Monte Carlo code MCNP. The unit has been realistically modelled: the cylindrical source capsule and its housing, the rectangular collimator system, both the primary and secondary jaws and the air gaps between the components. Different collimator openings, ranging from 5 x 5 cm2 to 20 x 20 cm2 (narrow and broad beams) at a source-surface distance equal to 80 cm have been used during the study. In the present work, we have calculated spectra as a function of field size. A study of the variation of the electron contamination of the 60Co beam has also been performed.