Khir Ashraf William, Swalen Marcel J, Segers Patrick, Verdonck Pascal, Pepper John R
Brunel Institute for Bioengineering and School of Engineering and Design, Brunel University, Uxbridge, Middlesex, England.
Artif Organs. 2006 Apr;30(4):308-12. doi: 10.1111/j.1525-1594.2006.00218.x.
The BCM (CardialCare, Minneapolis, MN, U.S.A.) is a pusher-plate pulsatile left ventricular assist device (LVAD) that is operated by counterpulsation pumps. The purpose of this work was to assess the fluid dynamics associated with operating the BCM in a mock circulation, and also to examine the similarities between hemodynamic parameters produced by this device in vitro and those produced by the left ventricle (LV) in vivo. The BCM was connected to a true size silicon rubber aorta and operated by an intra-aortic balloon pump. We examined the performance of the device at two system pressures (6.5 and 8 kPa); at three heart rates (60, 80, and 100 bpm); and at three pumping frequencies (1:1, 1:2, 1:3). Pressure and flow were measured in the upper descending aorta, and wave intensity analysis was used to calculate the peak intensity and energy of the compression and expansion waves. Pressure and flow waveforms produced by the BCM LVAD in vitro under different loading conditions were similar to those observed in vivo under similar loadings. Pusher-plate-type LVADs can produce compression and expansion waves similar to those generated by healthy LV in vivo.
BCM(美国明尼阿波利斯市CardialCare公司生产)是一种由反搏泵驱动的推板型搏动性左心室辅助装置(LVAD)。本研究的目的是评估在模拟循环中操作BCM时的流体动力学,并研究该装置在体外产生的血流动力学参数与左心室(LV)在体内产生的血流动力学参数之间的相似性。将BCM连接到真实尺寸的硅橡胶主动脉上,并通过主动脉内球囊泵进行操作。我们在两种系统压力(6.5和8 kPa)、三种心率(60、80和100次/分钟)以及三种泵血频率(1:1、1:2、1:3)下检查了该装置的性能。在上降主动脉中测量压力和流量,并使用波强度分析来计算压缩波和膨胀波的峰值强度和能量。BCM LVAD在体外不同负荷条件下产生的压力和流量波形与在体内相似负荷条件下观察到的波形相似。推板型LVAD可以产生与健康LV在体内产生的类似的压缩波和膨胀波。