Pradhan A S, Bakshi A K
Health, Safety and Environment Group, Bhabha Atomic Research Centre, Mumbai 400085, India.
Radiat Prot Dosimetry. 2006;119(1-4):276-9. doi: 10.1093/rpd/nci559. Epub 2006 Apr 27.
The study of the well separated low and high temperature glow peaks in CaSO4:Dy and CaF2:Tm was undertaken to develop comparative data on supralinearity of low and high temperature glow peaks for their use in the estimation of elapsed time of exposure and to analyse their responses to low and high-linear energy transfer (LET) radiation. In CaSO4:Dy, unlike its dosimetric peak, the structure of glow peaks at approximately 140 and 400 degrees C remains unchanged (peak position changes within 4 degrees C) with 60Co gamma-ray exposure up to 1 kGy air kerma. The glow peaks at 140 degrees C exhibited higher supralinearity than that of the peaks at 240 and 400 degrees C. In CaF2:Tm, 110 degrees C glow peak exhibited higher supralinearity and higher response to high-LET radiation as compared with 150 degrees C glow peak. No correlation between glow peak temperature and supralinearity or the LET response was observed.
对硫酸钙镝(CaSO4:Dy)和氟化钙铥(CaF2:Tm)中分离良好的低温和高温发光峰进行了研究,以获取关于低温和高温发光峰超线性的比较数据,用于估计暴露时间,并分析它们对低线性能量转移(LET)和高LET辐射的响应。在CaSO4:Dy中,与剂量测定峰不同,在高达1 kGy空气比释动能的60Coγ射线照射下,约140℃和400℃处的发光峰结构保持不变(峰位置在4℃内变化)。140℃处的发光峰比240℃和400℃处的峰表现出更高的超线性。在CaF2:Tm中,与150℃发光峰相比,110℃发光峰表现出更高的超线性和对高LET辐射的更高响应。未观察到发光峰温度与超线性或LET响应之间的相关性。