Miljanić S, Ranogajec-Komor M, Knezević Z, Stuhec M, Prokić M
Ruder Bosković Institute, Bijenicka 54, 10000 Zagreb, Croatia.
Radiat Prot Dosimetry. 2006;119(1-4):191-6. doi: 10.1093/rpd/nci557. Epub 2006 May 18.
Recently, two new types of 'tissue equivalent' thermoluminescent detectors (TLDs) have aroused attention: LiF:Mg,Cu,Na,Si and Li2B4O7:Cu,Ag,P. In this work the characteristics of both detectors were compared with the characteristics of the well-known type LiF:Mg,Ti detector, TLD-100. The following properties were investigated: the glow curve structures, relative sensitivity, batch homogeneity and uniformity, detection threshold, reproducibility of the response, linearity in the wide dose range and fading. Also, the energy dependence for medium and low energy X rays was determined in the range of mean energies between 33 and 116 keV. The results confirmed 'tissue equivalency' of both new types in the investigated range of photon energies. LiF:Mg,Cu,Na,Si detector has very high sensitivity (approximately 75 times higher than that of TLD-100) and is convenient for use in a very low range of doses. Li2B4O7:Cu,Ag,P detector shows some improvements in comparison with the previously prepared types of lithium borate. The most important is the five times higher sensitivity than that of TLD-100. This detector is also very promising, especially in medical dosimetry.
最近,两种新型的“组织等效”热释光探测器(TLD)引起了关注:LiF:Mg,Cu,Na,Si和Li2B4O7:Cu,Ag,P。在这项工作中,将这两种探测器的特性与著名的LiF:Mg,Ti探测器TLD-100的特性进行了比较。研究了以下特性:发光曲线结构、相对灵敏度、批次均匀性和一致性、探测阈值、响应的再现性、宽剂量范围内的线性以及衰退。此外,还测定了平均能量在33至116keV之间的中低能X射线的能量依赖性。结果证实了这两种新型探测器在研究的光子能量范围内具有“组织等效性”。LiF:Mg,Cu,Na,Si探测器具有非常高的灵敏度(约比TLD-100高75倍),便于在非常低的剂量范围内使用。Li2B4O7:Cu,Ag,P探测器与先前制备的硼酸锂类型相比有一些改进。最重要的是灵敏度比TLD-100高五倍。这种探测器也非常有前景,特别是在医学剂量测定方面。