Strauss Kathy A, O'Kane Jessica, Cook Janine Denis
Department of Epidemiology and Preventive Medicine, University of Maryland, Baltimore, Baltimore, MD 21201, USA.
Clin Biochem. 2006 Aug;39(8):826-32. doi: 10.1016/j.clinbiochem.2006.04.022. Epub 2006 May 19.
Novel comparison of CDT isoforms as determined by CE with an FDA-approved immunoassay kit.
Subjects (n=51) were categorized by drinking status based on AUDIT questionnaire responses. CDT isoform analyses by CE were compared to a commercially available, FDA-approved immunoassay. The analytical specificity of the immunoassay kit was assessed by analysis with IEF.
Because of the poor correlation between % CDT as measured by CE and the TIA immunoassay and between subject-reported drinking levels and results from the TIA assay, extraction column eluants from the immunoassays were analyzed by IEF for analytical specificity. % CDT by TIA included some trisialo-Tf, a non-CDT fraction, in the % CDT determination.
Total % CDT by CE, which separates all isoforms is more analytically specific than immunoassays because it does not include trisialo-Tf in the CDT calculation.
通过毛细管电泳(CE)测定的CDT同工型与美国食品药品监督管理局(FDA)批准的免疫分析试剂盒之间的新型比较。
根据酒精使用障碍识别测试(AUDIT)问卷的回答,将受试者(n = 51)按饮酒状况进行分类。将通过CE进行的CDT同工型分析与市售的、FDA批准的免疫分析进行比较。通过等电聚焦(IEF)分析评估免疫分析试剂盒的分析特异性。
由于CE测定的CDT百分比与TIA免疫分析之间以及受试者报告的饮酒水平与TIA分析结果之间的相关性较差,因此通过IEF分析免疫分析的萃取柱洗脱液的分析特异性。TIA测定的CDT百分比在CDT测定中包括一些三唾液酸转铁蛋白(trisialo-Tf),这是一种非CDT组分。
通过CE测定的总CDT百分比可分离所有同工型,在分析上比免疫分析更具特异性,因为它在CDT计算中不包括三唾液酸转铁蛋白。