Delta34S measurements of sulfur by multicollector inductively coupled plasma mass spectrometry.

作者信息

Clough Robert, Evans Peter, Catterick Tim, Evans E Hywel

机构信息

Speciation and Environmental Analysis Research Group, School of Earth Ocean and Environmental Sciences, University of Plymouth, Drake Circus, Plymouth, PL4 8AA, UK.

出版信息

Anal Chem. 2006 Sep 1;78(17):6126-32. doi: 10.1021/ac060875h.

Abstract

An accurate and precise method for the determination of delta34S measurements by multicollector inductively coupled plasma mass spectrometry has been developed. Full uncertainty budgets, taking into consideration all the uncertainties of the measurement process, have been calculated. The technique was evaluated by comparing measured values with a range of isotopically enriched sulfur solutions prepared by gravimetric addition of a 34S spike. The gravimetric and measured results exhibited a correlation of R2 >0.999. Repeat measurements were also made after adding Na (up to 420 microg g(-1)) and Ca (up to 400 microg g(-1)) salts to the sulfur standard. No significant deviations in the delta34S values were observed. The Russell correction expression (Ingle, C.; Sharp, B.; Horstwood, M.; Parrish, R.; Lewis, D. J. J. Anal. At. Spectrom. 2003, 18, 219) was used to correct for mass bias on the 34S/32S isotope amount ratio from the mass bias observed for the 30Si/28Si isotope amount ratio. Consistent compensation for instrumental mass bias was achieved. Resolution of the measured delta34S values was better than 1 per thousand after consideration of all uncertainty components. The technique was evaluated for practical applications by measurement of delta34S for a range of mineral waters by pneumatic nebulization sample introduction and the analysis of genuine and counterfeit pharmaceuticals using both laser ablation sample introduction and liquid chromatography. For the former two cases polyatomic interferences were resolved by operating the MC-ICPMS in medium resolution, while for the chromatographic analyses polyatomic interferences were minimized by the use of a membrane desolvator, allowing the instrument to be operated at a resolution of 400.

摘要

已开发出一种通过多接收电感耦合等离子体质谱法测定δ³⁴S测量值的准确且精确的方法。考虑到测量过程中的所有不确定性,计算了完整的不确定度预算。通过将测量值与一系列通过重量法添加³⁴S标样制备的同位素富集硫溶液进行比较,对该技术进行了评估。重量法和测量结果显示R²>0.999的相关性。在向硫标准中添加Na(高达420μg g⁻¹)和Ca(高达400μg g⁻¹)盐后,也进行了重复测量。未观察到δ³⁴S值有明显偏差。使用罗素校正表达式(英格尔,C.;夏普,B.;霍斯特伍德,M.;帕里什,R.;刘易斯,D. J. J. Anal. At. Spectrom. 2003, 18, 219)根据³⁰Si/²⁸Si同位素丰度比观察到的质量偏倚来校正³⁴S/³²S同位素丰度比的质量偏倚。实现了对仪器质量偏倚的一致补偿。考虑所有不确定度分量后,测量的δ³⁴S值的分辨率优于千分之一。通过使用气动雾化进样测量一系列矿泉水的δ³⁴S以及使用激光烧蚀进样和液相色谱分析正品和假冒药品,对该技术进行了实际应用评估。对于前两种情况,通过在中分辨率下操作MC-ICPMS来解决多原子干扰,而对于色谱分析,通过使用膜去溶剂器将多原子干扰降至最低,从而使仪器能够在400的分辨率下运行。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验