Usrey Molly M, Noble Philip C, Rudner Lanny J, Conditt Michael A, Birman Michael V, Santore Richard F, Mathis Kenneth B

机构信息

Institute of Orthopedic Research and Education, Houston, Texas, USA.

出版信息

J Arthroplasty. 2006 Sep;21(6 Suppl 2):65-71. doi: 10.1016/j.arth.2006.05.013.

Abstract

We explore the hypothesis that repetitive impingement increases wear of ultrahigh-molecular-weight polyethylene inserts. During revision total hip arthroplasty, 113 acetabular components were retrieved and examined for impingement, backside wear, and articular wear. Sixty percent of acetabular liners showed evidence of prosthetic impingement, with moderate or severe damage to the rim in 32%. Moderate or severe backside damage was present in 31%. Backside wear was found in 61% (22/36) of liners exhibiting impingement vs 16% (12/73) of liners without impingement (P < .0001). The average articular wear rate of liners with moderate/severe impingement damage was 159 +/- 42 mm(3)/y, compared with 70 +/- 21 mm(3)/y for liners with no/mild impingement damage (P = .02). Repetitive impingement frequently occurs after total hip arthroplasty and leads to a multitude of adverse events, including motion between the liner and the acetabular shell.

摘要

我们探讨了重复性撞击会增加超高分子量聚乙烯内衬磨损这一假说。在翻修全髋关节置换术中,取出113个髋臼组件,检查是否存在撞击、背面磨损和关节面磨损情况。60%的髋臼衬垫显示有假体撞击迹象,其中32%的边缘有中度或重度损伤。31%存在中度或重度背面损伤。在有撞击的衬垫中,61%(22/36)发现有背面磨损,而无撞击的衬垫中这一比例为16%(12/73)(P <.0001)。有中度/重度撞击损伤的衬垫平均关节面磨损率为159±42 mm³/年,而无/轻度撞击损伤的衬垫为70±21 mm³/年(P =.02)。重复性撞击在全髋关节置换术后经常发生,并会导致多种不良事件,包括衬垫与髋臼杯之间的活动。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验