Suppr超能文献

虹膜夹型有晶状体眼人工晶状体植入术的超声生物显微镜检查

Ultrasound biomicroscopy of iris-claw phakic intraocular lens implantation.

作者信息

Yu Ayong, Lin Zhende, Cai Xiaoyu, Chen Xiuqi, Luo Shuke, Li Yonghua

机构信息

Key Laboratory of Opthalmology of the Ministry of Education and Zhongshan Ophthalmic Center, Sun Yat-sen University, Guangzhou, 510060 China.

出版信息

Yan Ke Xue Bao. 2006 Mar;22(1):35-9.

Abstract

PURPOSE

To study in situ the intraocular position of iris-claw phakic intraocular lens (ICPIOL) in myopic eyes using ultrasound biomicroscopy (UBM).

METHODS

UBM echograms of the anterior segment were taken preoperatively and 62 to 115 days postoperatively in 6 eyes implanted with the Verisys ICPIOL (AMO). The echograms were assessed for the effect of the ICPIOL on iris tissue.

RESULTS

The preoperative distance between the corneal endothelium and the lens ranged from 2.96 to 3.09 mm, and the postoperative distance between the ICPIOL and the corneal endothelium, from 1.86 to 2.03 mm. The distance between the lens and the posterior surface of the ICPIOL ranged from 0.61 to 0.76 mm. The distance between the superior, inferior optic edge and the iris ranged from 0.49 to 1.00 mm, 0.21 to 0.51 mm respectively. The shortest distance between the ICPIOL haptics and the angle of anterior chamber ranged from 1.25 to 1.65 mm. The indentation of iris tissue by the ICPIOL haptics without pigmentary dispersion and distortion of posterior curvature of iris was observed.

CONCLUSION

Adequate space is maintained between the Verisyse myopic ICPIOL and the corneal endothelium, angle, and crystalline lens. Haptic indentation of the iris without pigment erosion and distortion of iris curvature is noted. The ICPIOL implanted in phakic eyes is a safe alternative for treatment of high myopia.

摘要

目的

使用超声生物显微镜(UBM)在原位研究虹膜夹型有晶状体眼人工晶状体(ICPIOL)在近视眼眼中的眼内位置。

方法

对6只植入Verisys ICPIOL(AMO)的眼睛在术前及术后62至115天拍摄眼前节的UBM超声图。评估超声图以观察ICPIOL对虹膜组织的影响。

结果

术前角膜内皮与晶状体之间的距离为2.96至3.09mm,术后ICPIOL与角膜内皮之间的距离为1.86至2.03mm。晶状体与ICPIOL后表面之间的距离为0.61至0.76mm。ICPIOL上缘、下缘光学边缘与虹膜之间的距离分别为0.49至1.00mm、0.21至0.51mm。ICPIOL襻与前房角之间最短距离为1.25至1.65mm。观察到ICPIOL襻对虹膜组织有压痕,但无色素播散及虹膜后表面曲率变形。

结论

Verisyse近视ICPIOL与角膜内皮、房角及晶状体之间保持了足够的空间。注意到虹膜襻压痕无色素侵蚀及虹膜曲率变形。植入有晶状体眼的ICPIOL是治疗高度近视的一种安全选择。