Torsional stability of intramedullary compression nails: tibial osteotomy model.

作者信息

Brown Nicholas A T, Bryan Nathaniel A, Stevens Peter M

机构信息

Department of Orthopaedics, University of Utah, 590 Wakara Way, Room A100, Salt Lake City, UT 84108, USA.

出版信息

Clin Biomech (Bristol). 2007 May;22(4):449-56. doi: 10.1016/j.clinbiomech.2006.11.009. Epub 2007 Jan 23.

Abstract

BACKGROUND

Studies comparing intramedullary compression nailing to conventional dynamized intramedullary nailing contend that better clinical outcomes of intramedullary compression techniques result from greater rotational stability of fracture or osteotomy sites. However, there appears to be no experimental evidence that rotational stability is improved with intramedullary nail compression. This study evaluated the effect of intramedullary nail compression on cadaveric and composite tibial fragment rotation above and below a mid-tibial transverse osteotomy.

METHODS

Twelve composite and four matched pairs of human cadaveric tibiae were randomized into either a compression group or a non-compressed dynamized group. A bi-axial servo-hydraulic testing machine applied 5Nm of internal and external torque (2Nm/s) under constant axial loads of 375N and 750N. Rotation along each tibia and at the osteotomy site was recorded from the trajectories of infrared emitting diodes rigidly attached along the tibial shaft.

FINDINGS

In comparison to dynamized nails, intramedullary nail compression significantly reduced the rotation at the osteotomy site in both human (8.9 degrees reduction; P=0.007) and composite tibiae (2.5 degrees reduction; P=0.039) under 750N of axial load. A similar result was noted for simulated half-body weight loads (375N) for both composite (3.0 degrees reduction; P=0.009) and human tibiae (10.5 degrees reduction; P=0.003).

INTERPRETATION

Intramedullary nail compression more effectively reduced tibial fragment rotation about a mid-tibial osteotomy than conventional dynamized intramedullary nails because intramedullary nails created more osteotomy site compression than the application of body weight to tibiae instrumented with dynamized intramedullary nails.