Lymperopoulou G, Papagiannis P, Angelopoulos A, Karaiskos P, Georgiou E, Baltas D
Nuclear and Particle Physics Section, Physics Department, University of Athens, Panepistimioupolis, Ilisia, 157 71, Athens, Greece.
Med Phys. 2006 Dec;33(12):4583-9. doi: 10.1118/1.2392408.
Monte Carlo simulation dosimetry is used to compare 169Yb to 192Ir for breast high dose rate (HDR) brachytherapy applications using multiple catheter implants. Results for bare point sources show that while 169Yb delivers a greater dose rate per unit air kerma strength at the radial distance range of interest to brachytherapy in homogeneous water phantoms, it suffers a greater dose rate deficit in missing scatter conditions relative to 192Ir. As a result of these two opposing factors, in the scatter conditions defined by the presence of the lung and the finite patient dimensions in breast brachytherapy the dose distributions calculated in a patient equivalent mathematical phantom by Monte Carlo simulations for the same implant of either 169Yb or 1921r commercially available sources are found comparable. Dose volume histogram results support that 169Yb could be at least as effective as 192Ir delivering the same dose to the lung and slightly reduced dose to the breast skin. The current treatment planning systems' approach of employing dosimetry data precalculated in a homogeneous water phantom of given shape and dimensions, however, is shown to notably overestimate the delivered dose distribution for 169Yb. Especially at the skin and the lung, the treatment planning system dose overestimation is on the order of 15%-30%. These findings do not undermine the potential of 169Yb HDR sources for breast brachytherapy relative to the most commonly used 192Ir HDR sources. They imply, however, that there could be a need for the amendment of dose calculation algorithms employed in clinical treatment planning of particular brachytherapy applications, especially for intermediate photon energy sources such as 169Yb.
蒙特卡罗模拟剂量测定法用于比较169镱和192铱在乳腺高剂量率(HDR)近距离放射治疗中的应用,采用多导管植入。裸点源的结果表明,虽然在均匀水模体中,对于近距离放射治疗感兴趣的径向距离范围内,169镱每单位空气比释动能强度的剂量率更高,但相对于192铱,在缺失散射条件下它的剂量率亏缺更大。由于这两个相反的因素,在乳腺近距离放射治疗中由肺的存在和有限的患者尺寸所定义的散射条件下,通过蒙特卡罗模拟在患者等效数学模体中计算得到的、使用市售的169镱或192铱源进行相同植入的剂量分布具有可比性。剂量体积直方图结果支持,169镱在向肺部输送相同剂量且向乳腺皮肤输送的剂量略有降低的情况下,至少与192铱一样有效。然而,目前治疗计划系统采用在给定形状和尺寸的均匀水模体中预先计算的剂量测定数据的方法,被证明会显著高估169镱的实际剂量分布。特别是在皮肤和肺部,治疗计划系统的剂量高估约为15% - 30%。这些发现并不削弱169镱高剂量率源相对于最常用的192铱高剂量率源在乳腺近距离放射治疗中的潜力。然而,这意味着可能需要修正特定近距离放射治疗应用临床治疗计划中所采用的剂量计算算法,尤其是对于诸如169镱这样的中等光子能量源。