Suppr超能文献

头颅侧位X线片剂量降低的可能性及其对临床诊断的影响。

Possibilities of dose reduction in lateral cephalometric radiographs and its effects on clinical diagnostics.

作者信息

Kaeppler G, Dietz K, Reinert S

机构信息

Zentraler Röntgenbereich, Zentrum für ZMK, Osianderstrasse 2-8, Universität Tübingen, D-72076 Tübingen, Germany.

出版信息

Dentomaxillofac Radiol. 2007 Jan;36(1):39-44. doi: 10.1259/dmfr/15761373.

Abstract

OBJECTIVES

The aim of the present study was to determine (1) the absorbed and the exit radiation doses for cephalometric exposures on a phantom head with various exposure settings and image receivers, and (2) the diagnostic image quality for various modalities assessed on cephalometric radiographs of patients.

METHODS

The dose measurements for lateral cephalometric radiographs were performed with a semiconductor detector, and also with thermoluminescent detectors and an Alderson phantom. Both the integral and the effective doses were determined. Two radiographs of each patient (n=119) were taken at two different times, one at a low tube potential setting, 75+/-5 kV, and one with a decreased dose. Film-screen systems with speed class 400 and one storage phosphor plate were used. Five observers assessed the radiographs for the visualization of six cephalometric reference points on a three-point scale with -1, 0 and 1. Twenty-seven image pairs were rescored to determine inter- and intrarater reliability. The statistical analysis was done using analysis of variance and Tukey's HSD (honestly significant difference) post hoc test.

RESULTS

Increasing the tube potential setting led to an average dose reduction to 83% (integral dose) or to 87% (effective dose). Instead of taking the radiograph at a low tube potential setting (75 kV), a dose reduction of about 15% was feasible at a high tube potential setting (90 kV). A significant difference in reference point visibility existed between film radiographs at low tube potential settings (mean score 0.984) and at high tube potential settings (90 kV, mean score 0.958). For the storage phosphor plates, there was no significant difference to the film-screen combinations at low tube potential and halved milliampere seconds settings. In the second assessment, there was a high degree of agreement (96.6%) compared with the first assessment (unadjusted for random agreement).

CONCLUSIONS

As there is only minimal dose reduction at increased tube potential settings, for a dose reduction, it seems to be more useful to use storage phosphor plates at unchanged tube potential and halved milliampere seconds settings compared with the film-screen combination.

摘要

目的

本研究的目的是确定:(1)在具有不同曝光设置和图像接收器的仿真头模型上进行头颅侧位片曝光时的吸收辐射剂量和出射辐射剂量;(2)在患者的头颅侧位片上评估的各种成像方式的诊断图像质量。

方法

使用半导体探测器、热释光探测器和阿尔德森仿真人体模型进行头颅侧位片的剂量测量。确定了积分剂量和有效剂量。对每位患者(n = 119)在两个不同时间拍摄两张X光片,一张在低管电压设置(75±5 kV)下拍摄,另一张采用降低剂量拍摄。使用了速度等级为400的屏-片系统和一块存储磷光板。五名观察者以-1、0和1的三点量表评估X光片上六个头颅测量参考点的可视化情况。对27对图像进行重新评分以确定观察者间和观察者内的可靠性。使用方差分析和Tukey's HSD(真实显著差异)事后检验进行统计分析。

结果

提高管电压设置导致平均剂量降低至83%(积分剂量)或87%(有效剂量)。与在低管电压设置(75 kV)下拍摄X光片相比,在高管电压设置(90 kV)下可行的剂量降低约15%。低管电压设置(平均得分0.984)和高管电压设置(90 kV,平均得分0.958)的胶片X光片在参考点可见性方面存在显著差异。对于存储磷光板,在低管电压和毫安秒设置减半的情况下,与屏-片组合相比没有显著差异。在第二次评估中,与第一次评估(未针对随机一致性进行调整)相比,一致性程度较高(96.6%)。

结论

由于在提高管电压设置时剂量降低幅度很小,因此为了降低剂量,与屏-片组合相比,在管电压不变且毫安秒设置减半的情况下使用存储磷光板似乎更有用。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验