Suppr超能文献

一种基于经皮导管的系统,用于测量适用于经胸除颤研究的电位梯度。

A percutaneous catheter-based system for the measurement of potential gradients applicable to the study of transthoracic defibrillation.

作者信息

Rosborough John P, Deno D Curtis, Walker Robert G, Niemann James T

机构信息

Department of Emergency Medicine, Harbor-UCLA Medical Center, Torrance, California 90509, USA.

出版信息

Pacing Clin Electrophysiol. 2007 Feb;30(2):166-74. doi: 10.1111/j.1540-8159.2007.00645.x.

Abstract

BACKGROUND

The local electric (E) field or potential gradient produced by a shock reliably predicts VF termination. In this study we evaluated a multiple electrode, catheter-based device for closed-chest 3D measurements of E field from transthoracic defibrillation shocks.

METHODS

Catheters with multiple electrodes on the tip were placed in intracardiac locations in anesthetized swine. An empirically derived calibration matrix and custom microprocessor was used to transform simultaneously measured voltages into orthogonal E field vector components. E fields produced in six intracardiac locations by 30 and 300 J shocks were compared in eight animals. Correlations were determined for measured current and E field at various shock strengths at two different transthoracic impedances in five additional animals. VF was induced in 12 animals and E field measured during defibrillation attempts.

RESULTS

The E field measurements resulting for 30 J transthoracic shocks were not significantly different among different intracardiac sites. At 300 J, however, significant differences were observed between sites with the greatest intensities recorded in the coronary sinus and right ventricle. Within animals, the variability of the measurement at each site was small, ranging from 2.8 +/- 1.6% to 5.7 +/- 4.5%. Significant correlations (P < 0.001) between measured E field and peak current were observed at native impedance (34 +/- 4 Omega, r = 0.81) and at adjusted impedance (76 +/- 4 Omega, r = 0.78) with transthoracic shocks of 200, 300, and 360 J. In VF studies, the probability of defibrillation was closely fit by a sigmoidal dose response curve in the coronary sinus E field with an approximate threshold of 4.7 V/cm with 50% defibrillation success at 9.3 V/cm.

CONCLUSIONS

The measured intracardiac E field variability within animals and at a specific site was small, exhibiting a median value of 5.1%, contrasted to median variabilities across animals of 5-11% suggesting the capacity of this measurement system to provide subject specific information on the distribution of E fields. The measured E field magnitudes across animals in the coronary sinus were linearly correlated with applied shock current with a very strong linear relation to effective shock voltage observed in vitro in a saline tank. When evaluated as a predictor of shock success, the observed values were consistent with previously reported critical fields. This technique may be of value in evaluating waveforms for transthoracic defibrillation as well as electrode size, placement, and composition.

摘要

背景

电击产生的局部电场(E)或电位梯度可可靠地预测室颤终止。在本研究中,我们评估了一种基于导管的多电极装置,用于在开胸状态下对经胸除颤电击产生的电场进行三维测量。

方法

将尖端带有多个电极的导管置于麻醉猪的心腔内位置。使用经验推导的校准矩阵和定制微处理器将同时测量的电压转换为正交电场矢量分量。在8只动物中比较了30 J和300 J电击在6个心腔内位置产生的电场。在另外5只动物中,测定了两种不同经胸阻抗下不同电击强度时测量电流与电场的相关性。在12只动物中诱发室颤,并在除颤尝试期间测量电场。

结果

30 J经胸电击产生的电场测量值在不同心腔内部位之间无显著差异。然而,在300 J时,在冠状窦和右心室记录到的强度最大的部位之间观察到显著差异。在动物体内,每个部位测量值的变异性较小,范围为2.8±1.6%至5.7±4.5%。在200、300和360 J经胸电击时,在自然阻抗(34±4Ω,r = 0.81)和调整阻抗(76±4Ω,r = 0.78)下,测量的电场与峰值电流之间存在显著相关性(P < 0.001)。在室颤研究中,冠状窦电场中的除颤概率紧密拟合为S形剂量反应曲线(除颤成功的近似阈值为4.7 V/cm,在9.3 V/cm时除颤成功率为50%)。

结论

在动物体内和特定部位测量的心腔内电场变异性较小,中位数为5.1%,而动物间的中位数变异性为5 - 11%,这表明该测量系统有能力提供关于电场分布的个体特异性信息。在冠状窦中,跨动物测量的电场大小与施加的电击电流呈线性相关,与在盐水槽中体外观察到的有效电击电压呈非常强的线性关系。当作为电击成功的预测指标进行评估时,观察值与先前报道的临界场一致。该技术在评估经胸除颤波形以及电极大小、放置和组成方面可能具有价值。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验