Department of Orthopedic Surgery, Beth Israel Medical Center and Albert Einstein College of Medicine, New York, NY, USA.
Orthopedics. 2007 Mar;30(3):218-21. doi: 10.3928/01477447-20070301-11.
During preparation for uncemented femoral arthroplasty, a phenomenon has been observed that indicates thinning of anterior metaphyseal bone to a critical level. Light can be seen from within the canal passing through the anterior cortex. This skylight sign alerts the surgeon that a cortical defect or fracture can occur on reaming, broaching, or component insertion. In 420 consecutive arthroplasties, a skylight sign was noted in 97 (23%) hips. In 5 of those hips an oval cortical defect was created and in 3 hips a fracture occurred during broaching or insertion. Loosening developed in 1 hip with fracture. No fractures or defects occurred in hips without a skylight sign. If a skylight sign is present, the femur is at risk and preventive measures should be taken.