Histologic and duplex comparison of the perclose and angio-seal percutaneous closure devices.

作者信息

Gargiulo Nicholas J, Veith Frank J, Ohki Takao, Scher Lawrence A, Berdejo George L, Lipsitz Evan C, Menegus Mark, Greenberg Mark

机构信息

Division of Vascular Surgery, Montefiore Medical Center, Albert Einstein College of Medicine, 111 East 20th Street, Bronx, NY 10467, USA.

出版信息

Vascular. 2007 Jan-Feb;15(1):24-9. doi: 10.2310/6670.2007.00004.

Abstract

The intravascular and extravascular effects of percutaneous closure devices have not been well studied. We assessed the performance and healing characteristics in dogs of two devices approved by the US Food and Drug Administration. Nine adult male dogs were anesthesized prior to percutaneous access of both femoral arteries with a 6F sheath. All dogs were systemically heparinized to an activated clotting time (ACT) > 250 seconds. Duplex sonography was performed preoperatively to measure vessel diameter and flow velocity. In each dog, one of two devices (Perclose, Abbot Laboratories, Abbott Park, IL or Angio-Seal, St. Jude Medical, St. Paul, MN) was randomly deployed into one of the two femoral arteries. The other device was deployed on the opposite side. Duplex sonography was repeated immediately after deployment and 28 days later to measure changes in vessel diameter and flow velocity. At 28 days, angiography was performed on both femoral arteries before they were removed for histologic evaluation. The time required to excise each vessel reflected the degree of scarring. Hemostasis time for the Angio-Seal device far surpassed the Perclose device (39 +/- 7 vs 0 minutes; p < .05). Vessel narrowing was observed only at 28 days after deployment of the Angio-Seal device (p < .05). Extensive extravascular scarring was observed with the Angio-Seal device, which resulted in a longer femoral artery dissection time and greater periadventitial scar thickness compared with the Perclose device (p < .05). When compared with the Perclose suture closure device, the Angio-Seal collagen plug closure device prolonged hemostasis time and produced greater vessel narrowing and periadventitial inflammation (extravascular scarring) in a canine model at 4 weeks.

摘要

经皮闭合装置的血管内和血管外效应尚未得到充分研究。我们评估了美国食品药品监督管理局批准的两种装置在犬类中的性能和愈合特性。9只成年雄性犬在经皮穿刺双侧股动脉并置入6F鞘管前进行麻醉。所有犬均全身肝素化,使活化凝血时间(ACT)>250秒。术前进行双功超声检查以测量血管直径和流速。在每只犬中,将两种装置之一(Perclose,雅培实验室,伊利诺伊州雅培公园;或Angio-Seal,圣犹达医疗公司,明尼苏达州圣保罗)随机置入两条股动脉中的一条。另一种装置则置入对侧。装置置入后立即及28天后重复进行双功超声检查,以测量血管直径和流速的变化。在28天时,对双侧股动脉进行血管造影,然后将其取出进行组织学评估。切除每条血管所需的时间反映了瘢痕形成的程度。Angio-Seal装置的止血时间远远超过Perclose装置(39±7分钟对0分钟;p<.05)。仅在Angio-Seal装置置入28天后观察到血管狭窄(p<.05)。与Perclose装置相比,Angio-Seal装置观察到广泛的血管外瘢痕形成,这导致股动脉解剖时间更长,外膜周围瘢痕厚度更大(p<.05)。与Perclose缝合闭合装置相比,在4周龄的犬模型中,Angio-Seal胶原塞闭合装置延长了止血时间,并导致更大的血管狭窄和外膜周围炎症(血管外瘢痕形成)。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验