Luszik-Bhadra M, Bolognese-Milsztajn T, Boschung M, Coeck M, Curzio G, Derdau D, d'Errico F, Fiechtner A, Kyllönen J-E, Lacoste V, Lievens B, Lindborg L, Lovefors Daun A, Reginatto M, Schuhmacher H, Tanner R, Vanhavere F
Physikalisch-Technische Bundesanstalt, D-38116 Braunschweig, Germany.
Radiat Prot Dosimetry. 2007;125(1-4):293-9. doi: 10.1093/rpd/ncm190. Epub 2007 Apr 22.
Within the EC project EVIDOS ('Evaluation of Individual Dosimetry in Mixed Neutron and Photon Radiation Fields'), different types of active neutron personal dosemeters (and some passive ones) were tested in workplace fields at nuclear installations in Europe. The results of the measurements which have been performed up to now are summarised and compared to our currently best estimates of the personal dose equivalent Hp(10). Under- and over-readings by more than a factor of two for the same dosemeter in different workplace fields indicate that in most cases the use of field-specific correction factors is required.
在欧盟委员会的EVIDOS项目(“混合中子与光子辐射场中个人剂量测定评估”)中,不同类型的有源中子个人剂量计(以及一些无源剂量计)在欧洲核设施的工作场所辐射场中进行了测试。现将截至目前已开展的测量结果进行总结,并与我们目前对个人剂量当量Hp(10)的最佳估计值进行比较。同一剂量计在不同工作场所辐射场中的读数低于或高于两倍以上,这表明在大多数情况下需要使用特定辐射场的校正因子。