Kort N P, van Raay J J A M, Thomassen B J W
Department of Orthopedics, Maasland Hospital, PO Box 5500, 6130 MB Sittard, The Netherlands.
Knee. 2007 Aug;14(4):280-3. doi: 10.1016/j.knee.2007.04.007. Epub 2007 May 31.
Use of an intramedullary rod is advised for the alignment of the femoral component of an Oxford phase-III prosthesis. There are users moving toward extramedullary alignment, which is merely an indicator of frustration with accuracy of intramedullary alignment. The results of our study with 10 cadaver femora demonstrate that use of a short and long intramedullary femoral rod may result in excessive flexion alignment error of the femoral component. Understanding of the extramedullary alignment possibility and experience with the visual alignment of the femoral drill guide is essential toward minimizing potential errors in the alignment of the femoral component.
建议使用髓内杆来对齐牛津三期假体的股骨部件。有些使用者正转向髓外对齐,这仅仅表明他们对髓内对齐的准确性感到沮丧。我们对10具尸体股骨的研究结果表明,使用短和长的股骨髓内杆可能会导致股骨部件出现过度的屈曲对齐误差。了解髓外对齐的可能性以及股骨钻孔导向器视觉对齐的经验对于将股骨部件对齐中的潜在误差降至最低至关重要。