Ozawa Soji, Morikawa Yasuhide, Oguma Junya, Kitagawa Yuko, Asada Hironori, Kitajima Masaki
Department of Surgery, School of Medicine, Keio University, Tokyo, Japan.
J Surg Res. 2008 Apr;145(2):266-71. doi: 10.1016/j.jss.2007.03.057. Epub 2007 Jun 11.
Intracorporeal suturing during endoscopic surgery is a difficult procedure. The curved needle is the most difficult to position properly within the jaws of the needle driver. The low friction suture materials exhibit good knot run-down, but at the expense of increased tendency for slipping. It is therefore necessary to invent a new needle and a new thread to facilitate intracorporeal suturing.
We made a new flat needle (F-needle) to facilitate grasping and a new reduced silicone-coated thread to prevent the knots from loosening. The time required to grasp the F-needle and the round needle (R-needle) by 10 surgeons was measured. The angle between the needle and the needle driver was measured. The minimum number of intertwining knots required for the knot to hold was counted, and the minimum number of superimposed knots that prevented the knots from coming undone when the testing machine attempted to release them was counted.
The F-needle was grasped in less time than the R-needle. The error in yaw angle with the F-needle was smaller than in the R-needle. The minimum number of knots needed to hold against the untying force with the reduced silicone-coated thread was less than with the conventional silicone-coated thread.
The new flat needle was grasped more accurately and quickly than the round needles, and the thread with reduced surface coating was less likely to loosen. Therefore, the combination of the new flat needle and thread with reduced surface coating is useful for endoscopic suturing.
内镜手术中的体内缝合是一项困难的操作。弯曲针在持针器钳口内最难正确定位。低摩擦缝线材料打结顺畅,但代价是滑脱倾向增加。因此,有必要发明一种新的针和新的线来便于体内缝合。
我们制作了一种便于抓取的新型扁平针(F针)和一种防止结松开的新型减涂层硅胶线。测量了10名外科医生抓取F针和圆形针(R针)所需的时间。测量了针与持针器之间的角度。计算使结固定所需的最小缠绕结数,并计算在测试机器试图解开结时防止结松开所需的最小叠加结数。
抓取F针的时间比R针短。F针的偏航角误差比R针小。使用减涂层硅胶线抵抗解结力所需的最小结数比传统硅胶线少。
新型扁平针比圆形针抓取更准确、更快,表面涂层减少的线不太容易松开。因此,新型扁平针与表面涂层减少的线的组合对内镜缝合很有用。