Suppr超能文献

活动轴承膝关节中聚乙烯磨损的定量分析:前后滑动(APG)和旋转平台(RP)低接触应力(LCS)膝关节的翻修分析

Quantification of polyethylene degradation in mobile bearing knees: a retrieval analysis of the Anterior-Posterior-Glide (APG) and Rotating Platform (RP) Low Contact Stress (LCS) knee.

作者信息

Kop Alan M, Swarts Eric

机构信息

Department of Medical Engineering and Physics, Royal Perth Hospital, Perth, Western Australia.

出版信息

Acta Orthop. 2007 Jun;78(3):364-70. doi: 10.1080/17453670710013942.

Abstract

BACKGROUND

Routine qualitative observations of more than 850 polyethylene fixed and mobile bearings at our institution have noted minimal wear of mobile bearings. The APG mobile bearing is the most recent design variant of the LCS knee, allows multi-directional movement at the tibiofemoral articulation, and is posterior cruciate sparing. Even though it is difficult to perform, quantitative wear measurement is important in determining the likely longevity of new arthroplasty devices, and is especially relevant because of increasing numbers of new mobile bearing designs.

PATIENTS AND METHODS

We analyzed 10 retrieved APG and 7 retrieved RP tibial bearings (De Puy) with a mean implantation period of 33 (9-70) months. We used coordinate-measuring techniques to quantitatively determine linear penetration, and optical and scanning electron microscopy to assess wear mechanisms qualitatively.

RESULTS

The mean total volume loss (superior and inferior articulations) of the APG and RP designs was 85 mm((3))/year and 77 mm((3))/year, respectively. Burnishing was the predominant wear mechanism, and to a lesser extent scratching, abrasion and pitting. Multidirectional scratching and abrasion were noted on the APG inserts inferiorly, whereas there was circumferential scratching on the RP inserts.

INTERPRETATION

Our short-term results for the APG and RP mobile bearing designs are similar and compare more than favorably with reported values for fixed-bearing designs. However, increased backside wear due to multidirectional movement may predispose the APG design to greater wear in the long term.

摘要

背景

在我们机构对850多个聚乙烯固定和活动轴承进行的常规定性观察发现,活动轴承磨损极小。APG活动轴承是LCS膝关节的最新设计变体,允许在胫股关节处进行多方向运动,并且保留后交叉韧带。尽管难以进行,但定量磨损测量对于确定新型关节置换装置的可能使用寿命很重要,而且由于新型活动轴承设计数量不断增加,这一点尤为重要。

患者和方法

我们分析了10个取出的APG胫骨轴承和7个取出的RP胫骨轴承(Depuy公司),平均植入时间为33(9 - 70)个月。我们使用坐标测量技术定量确定线性穿透,并使用光学和扫描电子显微镜定性评估磨损机制。

结果

APG和RP设计的平均总体积损失(上关节面和下关节面)分别为每年85立方毫米和每年77立方毫米。研磨是主要的磨损机制,其次是刮擦、磨损和点蚀。在APG垫片下方观察到多方向刮擦和磨损,而RP垫片上有圆周刮擦。

解读

我们对APG和RP活动轴承设计的短期结果相似,并且比报道的固定轴承设计值更具优势。然而,由于多方向运动导致的背面磨损增加可能使APG设计在长期内更容易出现更大的磨损。