Kunz Martin, Caldwell Wendel A, Miyagi Lowell, Wenk Hans-Rudolf
Advanced Light Source, Lawrence Berkeley Laboratory, Berkeley, CA 94720, USA.
Rev Sci Instrum. 2007 Jun;78(6):063907. doi: 10.1063/1.2749443.
We report a first combination of diamond anvil cell radial x-ray diffraction with in situ laser heating. The laser-heating setup of ALS beamline 12.2.2 was modified to allow one-sided heating of a sample in a diamond anvil cell with an 80 W yttrium lithium fluoride laser while probing the sample with radial x-ray diffraction. The diamond anvil cell is placed with its compressional axis vertical, and perpendicular to the beam. The laser beam is focused onto the sample from the top while the sample is probed with hard x-rays through an x-ray transparent boron-epoxy gasket. The temperature response of preferred orientation of (Fe,Mg)O is probed as a test experiment. Recrystallization was observed above 1500 K, accompanied by a decrease in stress.
我们报道了金刚石对顶砧径向X射线衍射与原位激光加热的首次结合。对先进光源(ALS)12.2.2光束线的激光加热装置进行了改进,以便在利用80W氟化钇锂激光对金刚石对顶砧中的样品进行单面加热的同时,用径向X射线衍射对样品进行探测。金刚石对顶砧的压缩轴垂直放置且与光束垂直。激光束从顶部聚焦到样品上,同时通过X射线透明的硼环氧树脂垫片用硬X射线对样品进行探测。作为一个测试实验,对(铁,镁)氧化物择优取向的温度响应进行了探测。在1500K以上观察到了再结晶现象,同时应力降低。