Hayashi M, Kaku D, Watanabe Y, Sagara K
Department of Advanced Energy Engineering Science, Kyushu University, Kasuga, Fukuoka 816-8580, Japan.
Radiat Prot Dosimetry. 2007;126(1-4):376-9. doi: 10.1093/rpd/ncm078. Epub 2007 Jul 6.
Characteristics of a (6)Li-loaded neutron coincidence spectrometer were investigated from both measurements and Monte Carlo simulations. The spectrometer consists of three (6)Li-glass scintillators embedded in a liquid organic scintillator BC-501A, which can detect selectively neutrons that deposit the total energy in the BC-501A using a coincidence signal generated from the capture event of thermalised neutrons in the (6)Li-glass scintillators. The relative efficiency and the energy response were measured using 4.7, 7.2 and 9.0 MeV monoenergetic neutrons. The measured ones were compared with the Monte Carlo calculations performed by combining the neutron transport code PHITS and the scintillator response calculation code SCINFUL. The experimental light output spectra were in good agreement with the calculated ones in shape. The energy dependence of the detection efficiency was reproduced by the calculation. The response matrices for 1-10 MeV neutrons were finally obtained.
通过测量和蒙特卡罗模拟研究了一种装载(6)Li的中子符合谱仪的特性。该谱仪由嵌入液体有机闪烁体BC - 501A中的三个(6)Li玻璃闪烁体组成,它可以利用(6)Li玻璃闪烁体中热化中子俘获事件产生的符合信号,选择性地探测在BC - 501A中沉积全部能量的中子。使用4.7、7.2和9.0 MeV的单能中子测量了相对效率和能量响应。将测量结果与通过结合中子输运代码PHITS和闪烁体响应计算代码SCINFUL进行的蒙特卡罗计算结果进行了比较。实验光输出谱在形状上与计算结果吻合良好。计算再现了探测效率的能量依赖性。最终获得了1 - 10 MeV中子的响应矩阵。