Baños-Capilla M C, García M A, Bea J, Pla C, Larrea L, López E
Department of Medical Physics, Radiation Oncology Hospital Virgen del Consuelo, Callosa de Ensarria 12-Valencia, Valencia 46007, Spain.
Med Phys. 2007 Jun;34(6):1911-7. doi: 10.1118/1.2732031.
The quality of dosimetry in radiotherapy treatment requires the accurate delimitation of the gross tumor volume. This can be achieved by complementing the anatomical detail provided by CT images through fusion with other imaging modalities that provide additional metabolic and physiological information. Therefore, use of multiple imaging modalities for radiotherapy treatment planning requires an accurate image registration method. This work describes tests carried out on a Discovery LS positron emission/computed tomography (PET/CT) system by General Electric Medical Systems (GEMS), for its later use to obtain images to delimit the target in radiotherapy treatment. Several phantoms have been used to verify image correlation, in combination with fiducial markers, which were used as a system of external landmarks. We analyzed the geometrical accuracy of two different fusion methods with the images obtained with these phantoms. We first studied the fusion method used by the PET/CT system by GEMS (hardware fusion) on the basis that there is satisfactory coincidence between the reconstruction centers in CT and PET systems; and secondly the fiducial fusion, a registration method, by means of least-squares fitting algorithm of a landmark points system. The study concluded with the verification of the centroid position of some phantom components in both imaging modalities. Centroids were estimated through a calculation similar to center-of-mass, weighted by the value of the CT number and the uptake intensity in PET. The mean deviations found for the hardware fusion method were: deltax/ +/-sigma = 3.3 mm +/- 1.0 mm and /deltax/ +/-sigma = 3.6 mm +/- 1.0 mm. These values were substantially improved upon applying fiducial fusion based on external landmark points: /deltax/ +/-sigma = 0.7 mm +/- 0.8 mm and /deltax/ +/-sigma = 0.3 mm 1.7 mm. We also noted that differences found for each of the fusion methods were similar for both the axial and helical CT image acquisition protocols.
放射治疗中的剂量测定质量要求准确界定大体肿瘤体积。这可以通过将CT图像提供的解剖细节与其他能提供额外代谢和生理信息的成像模态相融合来实现。因此,在放射治疗计划中使用多种成像模态需要一种精确的图像配准方法。本文描述了对通用电气医疗系统(GEMS)的Discovery LS正电子发射/计算机断层扫描(PET/CT)系统进行的测试,以便后续用于获取界定放射治疗靶区的图像。结合基准标记使用了多个模体来验证图像相关性,基准标记用作外部标志物系统。我们分析了使用这些模体获得的图像与两种不同融合方法的几何精度。我们首先研究了GEMS的PET/CT系统所使用的融合方法(硬件融合),前提是CT系统和PET系统的重建中心之间有令人满意的重合度;其次研究了基准融合,这是一种通过标志物点系统的最小二乘拟合算法的配准方法。该研究最后验证了两种成像模态中一些模体组件的质心位置。质心是通过类似于质心的计算来估计的,由CT值和PET中的摄取强度加权。硬件融合方法的平均偏差为:deltax / +/-sigma = 3.3毫米 +/- 1.0毫米和 /deltax/ +/-sigma = 3.6毫米 +/- 1.0毫米。基于外部标志物点应用基准融合后,这些值有了显著改善:/deltax/ +/-sigma = 0.7毫米 +/- 0.8毫米和 /deltax/ +/-sigma = 0.3毫米 1.7毫米。我们还注意到对于轴向和螺旋CT图像采集协议,两种融合方法的差异相似。