Lankester Benedict J A, Sabri Omar, Gheduzzi Sabina, Stoney James D, Miles Anthony W, Bannister Gordon C
Avon Orthopaedic Centre, Southmead Hospital, Westbury-on-Trym, Bristol BS10 5NB, United Kingdom.
J Arthroplasty. 2007 Aug;22(5):738-44. doi: 10.1016/j.arth.2006.09.016.
A model was developed to assess the effect of an acetabular flange on pressure within different zones of the cement mantle during insertion. Two prosthetic designs were assessed in 3 different sizes. Flanged components produced significantly higher mean pressures than unflanged ones (P < .01). The effect of a flange was more pronounced at the rim than at the pole. Delayed insertion resulted in a further significant rise in mean pressure (P < .01), but this did not compensate for the lack of a flange. This experimental model supports the use of a flange to increase pressure within the cement mantle on component insertion. The beneficial effect is more marked in the area that is most likely to show deficiency in the cement-bone interface on postoperative radiographs.
开发了一种模型,以评估髋臼凸缘在假体植入过程中对骨水泥壳不同区域内压力的影响。对两种不同的假体设计进行了三种不同尺寸的评估。带凸缘的部件产生的平均压力明显高于无凸缘的部件(P <.01)。凸缘在边缘处的作用比在极点处更明显。延迟植入导致平均压力进一步显著升高(P <.01),但这并不能弥补凸缘的缺失。该实验模型支持使用凸缘来增加假体植入时骨水泥壳内的压力。在术后X线片上最有可能显示骨水泥-骨界面缺陷的区域,这种有益效果更为明显。