Hu Wenling, Hu Bin, Jiang Zucheng
Department of Chemistry, Wuhan 430072, PR China.
Anal Chim Acta. 2006 Jul 14;572(1):55-62. doi: 10.1016/j.aca.2006.05.024. Epub 2006 May 16.
In this paper, an ordered mesoporous alumina coating was prepared and applied to capillary microextraction (CME) of trace Co, Ni and Cd for the first time. The coated capillary was used for on-line coupling CME with inductively plasma mass spectrometry (ICP-MS) for the determination of trace metals of Co, Ni and Cd. The porous structure of Al2O3 coating was examined by SEM and TEM. The effects of the extraction parameters including pH, sample flow rate and volume, elution solution and interfering ions on the recovery of analytes have been investigated and optimized. Under the optimized conditions, the limits of detection were 0.33, 1.5 and 1.4 ng L(-1) for Co, Ni and Cd, respectively, with a preconcentration factor of 10 times. The precisions for all investigated elements were 2.7, 4.1 and 2.5% (c=0.05 ng L(-1), n=7), for Co, Ni and Cd, respectively, and the sample frequency was 8 h(-1). The proposed method was successfully applied for the analysis of trace metals in water, rice and urine samples with the recovery of 94-105%. In order to validate the proposed method, two certified reference materials of GBW 0913 human urine and NIES No.10-b rice flour were analyzed, and the determination values are in good agreement with the certified values. The ordered mesoporous Al2O3 coated capillary can be used more than 20 times without decreasing the extraction efficiency.
本文首次制备了有序介孔氧化铝涂层,并将其应用于痕量钴、镍和镉的毛细管微萃取(CME)。将涂覆的毛细管用于CME与电感耦合等离子体质谱(ICP-MS)的在线联用,以测定钴、镍和镉等痕量金属。通过扫描电子显微镜(SEM)和透射电子显微镜(TEM)对Al2O3涂层的多孔结构进行了研究。考察并优化了萃取参数(包括pH值、样品流速和体积、洗脱溶液以及干扰离子)对分析物回收率的影响。在优化条件下,钴、镍和镉的检测限分别为0.33、1.5和1.4 ng L⁻¹,预富集倍数为10倍。所有研究元素的精密度,钴、镍和镉分别为2.7%、4.1%和2.5%(c = 0.05 ng L⁻¹,n = 7),进样频率为8 h⁻¹。所提出的方法成功应用于水、大米和尿液样品中痕量金属的分析,回收率为94% - 105%。为验证所提出的方法,分析了两种有证标准物质GBW 0913人尿和NIES No.10 - b米粉,测定值与认定值吻合良好。有序介孔Al₂O₃涂覆的毛细管可使用20多次而不降低萃取效率。