Xu Qianyi, Hamilton Russell J, Schowengerdt Robert A, Jiang Steve B
Department of Electrical and Computer Engineering, University of Arizona, Tucson, AZ 85721, USA.
Phys Med Biol. 2007 Sep 7;52(17):5277-93. doi: 10.1088/0031-9155/52/17/012. Epub 2007 Aug 16.
A dynamic multi-leaf collimator (DMLC) can be used to track a moving target during radiotherapy. One of the major benefits for DMLC tumor tracking is that, in addition to the compensation for tumor translational motion, DMLC can also change the aperture shape to conform to a deforming tumor projection in the beam's eye view. This paper presents a method that can track a deforming lung tumor in fluoroscopic video using active shape models (ASM) (Cootes et al 1995 Comput. Vis. Image Underst. 61 38-59). The method was evaluated by comparing tracking results against tumor projection contours manually edited by an expert observer. The evaluation shows the feasibility of using this method for precise tracking of lung tumors with deformation, which is important for DMLC-based real-time tumor tracking.
动态多叶准直器(DMLC)可用于在放射治疗期间跟踪移动目标。DMLC肿瘤跟踪的一个主要优点是,除了补偿肿瘤的平移运动外,DMLC还可以改变孔径形状,以在射野视角下符合变形肿瘤的投影。本文提出了一种使用主动形状模型(ASM)(Cootes等人,1995年,《计算机视觉与图像理解》61卷,38 - 59页)在荧光透视视频中跟踪变形肺肿瘤的方法。通过将跟踪结果与专家观察者手动编辑的肿瘤投影轮廓进行比较来评估该方法。评估表明,使用该方法精确跟踪变形肺肿瘤是可行的,这对于基于DMLC的实时肿瘤跟踪很重要。