Abstract

OBJECTIVE

In this study, we determined the assay performance criteria necessary to produce acceptable results for >or=98% of neonate bilirubin samples collected by capillary heel-stick.

STUDY DESIGN AND METHODS

We determined serum free hemoglobin levels in 151 heel-stick serum samples to determine the hemolysis level. We then tested the effect of hemolysis on total bilirubin levels determined by four commercially available assays.

RESULTS

The mean level of serum free hemoglobin was 1.62 g/L. Of the serum total bilirubin assays tested, the Total Bilirubin Special (Roche Diagnostics) and the TBILI (Roche Diagnostics) reagents did not show significant interference at the concentrations of free hemoglobin observed in >or=98% of heel-stick samples. The Vitros Bu/Bc slide (Ortho-Clinical Diagnostics) showed significant interference only at normal bilirubin concentrations; while the Bilirubin DPD reagent (Amresco Inc.) showed significant interference starting at hemoglobin concentrations of 1.0 g/L.

CONCLUSIONS

Bilirubin assays that are not sensitive to approximately 6 g/L free hemoglobin should provide accurate results for most samples obtained via capillary heel-stick. Of the four assays tested, the Bilirubin DPD reagent (Amresco Inc.) was the most susceptible to the presence of free hemoglobin and will result in a higher rejection rate of neonate capillary heel-stick samples.

摘要

目的

在本研究中,我们确定了对于通过足跟毛细血管采血收集的≥98%的新生儿胆红素样本产生可接受结果所需的检测性能标准。

研究设计与方法

我们测定了151份足跟采血血清样本中的血清游离血红蛋白水平以确定溶血程度。然后我们测试了溶血对四种市售检测方法测定总胆红素水平的影响。

结果

血清游离血红蛋白的平均水平为1.62 g/L。在所测试的血清总胆红素检测方法中,总胆红素专用检测法(罗氏诊断)和TBILI检测法(罗氏诊断)在≥98%的足跟采血样本中观察到的游离血红蛋白浓度下未显示出显著干扰。Vitros Bu/Bc玻片(奥森多临床诊断公司)仅在正常胆红素浓度下显示出显著干扰;而胆红素DPD试剂(爱默斯公司)在血红蛋白浓度达到1.0 g/L时开始显示出显著干扰。

结论

对约6 g/L游离血红蛋白不敏感的胆红素检测方法应为大多数通过足跟毛细血管采血获得的样本提供准确结果。在所测试的四种检测方法中,胆红素DPD试剂(爱默斯公司)对游离血红蛋白的存在最敏感,这将导致新生儿足跟毛细血管采血样本的拒收率更高。