Development of a continuous protein and polysaccharide measurement method by sequential injection analysis for application in membrane bioreactor systems.

作者信息

Mehrez R, Ernst M, Jekel M

机构信息

Department of Water Quality Control, Technical University of Berlin, Germany.

出版信息

Water Sci Technol. 2007;56(6):163-71. doi: 10.2166/wst.2007.645.

Abstract

Polysaccharides and proteins are the main components of extracellular polymeric substances, which are considered to give a major contribution to the overall fouling potential of the MBR system. This work focuses on the automation of spectrophotometrical assays for protein and polysaccharide determination for on-line measurement in MBR. Sequential Injection Analysis (SIA) was identified as more suitable for continuous measurements. In screening tests the selection of an appropriate assay for the successful automation by means of SIA was carried out. Lowry Assay and Dubois Assay were chosen for protein and polysaccharide determination. Lowry Assay could be successfully automated by means of SIA. The automated method shows a lower sensitivity than the manual Lowry method, what can be contributed to specific characteristic of SIA. The confidence limit was calculated to 2.8 mg/L. The automated method shows in validation tests good correspondence with protein concentrations measured with the manual assay (R(2)=0.971). The successful establishment of the Dubois Assay for polysaccharides failed so far due to sensitivity problems.

摘要

多糖和蛋白质是细胞外聚合物的主要成分,被认为对膜生物反应器(MBR)系统的整体污染潜力有重大贡献。这项工作的重点是用于MBR在线测量的蛋白质和多糖分光光度测定法的自动化。顺序注射分析(SIA)被认为更适合连续测量。在筛选试验中,通过SIA成功实现自动化的合适测定法被选定。选择了Lowry法和Dubois法分别用于蛋白质和多糖的测定。Lowry法可通过SIA成功实现自动化。自动化方法的灵敏度低于手动Lowry法,这可能归因于SIA的特定特性。计算出的置信限为2.8 mg/L。在验证试验中,自动化方法与手动测定法测得的蛋白质浓度具有良好的相关性(R² = 0.971)。到目前为止,由于灵敏度问题,用于多糖测定的Dubois法尚未成功建立。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验