Jalkanen Ville, Andersson Britt M, Bergh Anders, Ljungberg Börje, Lindahl Olof A
Department of Applied Physics and Electronics, Umeå University, SE-90187 Umeå, Sweden.
Physiol Meas. 2007 Oct;28(10):1267-81. doi: 10.1088/0967-3334/28/10/011. Epub 2007 Sep 28.
In recent years, tactile sensors based on piezoelectric resonance sensor technology have been used for medical diagnosis where the sensor's stiffness-measuring properties can reflect tissue pathology. The change in the frequency of the resonating system and the change in force when contact is made with tissue are used as a stiffness parameter. Earlier stiffness measurements of prostate tissue in vitro demonstrate variations related to tissue composition. In this study, measured stiffness from two human prostate specimens was compared to histological composition of prostate tissue below and around the measurement points. Tissue stiffness was measured with the resonance sensor system. Tissue composition was measured at four different depths in the tissue specimen using a microscopic-image-based morphometrical method. With this method, the proportion of tissue types was determined at the points of intersections in a circular grid on the images representing each measurement point. Numerical values were used for weighting the tissue proportions at different depths in the tissue specimen. For an impression depth of 1.0 mm, the sensing depth in this study was estimated to be 3.5-5.5 mm. Stiffness variations due to horizontal tissue variations were investigated by studying the dependence of the size of the circular grid area relative to the contact area of the sensor tip. The sensing area (grid radius) was estimated to be larger than the contact area (contact radius) between the sensor tip and the tissue. Thus, the sensor tip registers spatial variations in prostate tissue histology, both directly below and lateral to the tip itself. These findings indicate that tumours around the sensor tip could be detected, which in turn supports the idea of a future resonance-sensor-based clinical device for detecting tumours and for guiding biopsies.
近年来,基于压电共振传感器技术的触觉传感器已被用于医学诊断,其中传感器的刚度测量特性可以反映组织病理学。共振系统频率的变化以及与组织接触时的力的变化被用作刚度参数。早期对前列腺组织的体外刚度测量表明其与组织组成有关。在本研究中,将两个人类前列腺标本的测量刚度与测量点下方及周围前列腺组织的组织学组成进行了比较。使用共振传感器系统测量组织刚度。使用基于微观图像的形态计量学方法在组织标本的四个不同深度测量组织组成。通过这种方法,在代表每个测量点的图像上的圆形网格的交点处确定组织类型的比例。使用数值对组织标本不同深度处的组织比例进行加权。对于1.0毫米的压痕深度,本研究中的传感深度估计为3.5 - 5.5毫米。通过研究圆形网格区域大小相对于传感器尖端接触面积的依赖性,研究了水平组织变化引起的刚度变化。传感区域(网格半径)估计大于传感器尖端与组织之间的接触面积(接触半径)。因此,传感器尖端记录了前列腺组织组织学的空间变化,既包括尖端正下方的变化,也包括尖端侧面的变化。这些发现表明可以检测到传感器尖端周围的肿瘤,这反过来支持了未来基于共振传感器的临床设备用于检测肿瘤和引导活检的想法。