采用无孔涂层髋臼部件(AcSys希勒杯)的非骨水泥型全髋关节置换术效果不佳。

Poor results of the cementless total hip arthroplasty with a nonporous coated acetabular component: AcSys Shearer Cup.

作者信息

Ikema Masahide, Oyama Masamizu, Kita Atsushi, Funayama Kanichi

机构信息

Department of Orthopaedic Surgery, Japanese Red Cross Sendai Hospital, 2-43-3 Yagiyama-honchou, Taihaku-ku, Sendai, Miyagi, 982-8501, Japan.

出版信息

J Orthop Sci. 2007 Sep;12(5):466-70. doi: 10.1007/s00776-007-1161-y. Epub 2007 Sep 28.

Abstract

BACKGROUND

This study followed patients for a minimum of 7 years after primary total hip arthroplasty using cementless acetabular components and evaluated their outcomes.

METHODS

We followed 73 patients (75 hips), who had undergone total hip arthroplasty with cementless nonporous coated acetabular components (3M AcSys Shearer Cup) for a mean of 9.8 years (range 7-13 years). There were 61 women and 12 men with a mean age of 53 years (range 27-69 years) at surgery. The diagnosis was primary osteoarthritis in 9 hips, osteoarthritis secondary to developmental dysplasia in 58 hips, osteonecrosis of the femoral head in 6 hips, and rheumatoid arthritis in 2 hips.

RESULTS

Three cups were revised because of aseptic loosening, and one cup was revised following removal of the prosthesis due to deep infection. Radiographic loosening was observed in 22 hips at the latest follow-up. The survival rate at 10 years was 94.7% with revision as the endpoint and 72% with radiographic loosening as the endpoint. The Merle d'Aubigné and Postel hip score showed significant improvement postoperatively and was maintained well even in cases showing radiographic loosening.

CONCLUSIONS

The intermediate radiological results with the AcSys Shearer Cup were unsatisfactory because of the high loosening rate, although the revision rate was low. The nonporous outer surface and the poor fixation mechanism between the metal shell and liner may have contributed to the high failure rate. Regular radiological review is recommended when this cup is used because early loosening is often painless.

摘要

背景

本研究对使用非骨水泥髋臼组件进行初次全髋关节置换术后的患者进行了至少7年的随访,并评估了其预后情况。

方法

我们随访了73例患者(75髋),这些患者接受了使用非骨水泥无孔涂层髋臼组件(3M AcSys Shearer杯)的全髋关节置换术,平均随访时间为9.8年(范围7 - 13年)。手术时,有61名女性和12名男性,平均年龄53岁(范围27 - 69岁)。诊断为原发性骨关节炎9髋,发育性髋关节发育不良继发骨关节炎58髋,股骨头坏死6髋,类风湿关节炎2髋。

结果

3个髋臼杯因无菌性松动而翻修,1个髋臼杯因深部感染在取出假体后进行了翻修。在最近一次随访时,22髋出现影像学松动。以翻修为终点,10年生存率为94.7%;以影像学松动为终点,10年生存率为72%。Merle d'Aubigné和Postel髋关节评分术后有显著改善,即使在出现影像学松动的病例中也保持良好。

结论

AcSys Shearer杯的中期影像学结果不尽人意,尽管翻修率较低,但松动率较高。无孔外表面以及金属杯与内衬之间欠佳的固定机制可能导致了较高的失败率。使用该髋臼杯时建议定期进行影像学检查评估,因为早期松动通常无痛。