Dolgov Arseny B, Zane Regan
ECE Dept., Colorado Univ., Boulder, CO.
Conf Proc IEEE Eng Med Biol Soc. 2006;2006:2067-70. doi: 10.1109/IEMBS.2006.259564.
Long-term, low duty cycle monitoring of patients with a variety of disabilities or health concerns is often required. In this paper, we discuss the design considerations and implementation of an ultra-low power wireless medical sensor platform, suitable for a wide range of medical and sports applications. A hardware demonstration prototype based on readily available components is presented with sensors for 3-axis acceleration, temperature and galvanic skin response. Detailed power measurements and operation results are shown, demonstrating a sensor life span of more than 10 years on a single 200 mAh lithium watch battery using low current standby techniques with an average power of less than 5 muW and a 10 second sample interval.
通常需要对患有各种残疾或健康问题的患者进行长期、低占空比监测。在本文中,我们讨论了一种适用于广泛医疗和运动应用的超低功耗无线医疗传感器平台的设计考量与实现。展示了一个基于现成组件的硬件演示原型,其配备了用于三轴加速度、温度和皮肤电反应的传感器。给出了详细的功耗测量结果和运行结果,表明使用低电流待机技术,在单个200毫安时的锂手表电池上,以小于5微瓦的平均功率和10秒的采样间隔,传感器寿命超过10年。