Tang Liang, Hwang Gyo-Seung, Yang Lin, Lin Shien-Fong
Dept. of Medicine, Cedars-Sinai Med. Center, Los Angeles, CA 90048, USA.
Conf Proc IEEE Eng Med Biol Soc. 2006;2006:4362-5. doi: 10.1109/IEMBS.2006.260036.
Ventricular fibrillation (VF) can be synchronized with a novel synchronized pacing technique (SyncP) using low-energy pacing pulses, which causes pace termination of VF. Synchronized pacing (SyncP) is defined as optical recording guided real-time detection and stimulation of spatiotemporal excitable gaps. In this paper, we investigate the effect of post-shock SyncP strategy on improvement of defibrillation efficacy. After a near-threshold defibrillation shock, when the reference site detected the earliest activation of the reinitiated VF, a 5-mA electric stimulus was delivered from the post-shock pacing electrode to depolarize the excitable gap. This area of wavefront synchronization may lead to a change in the timing of VF propagation, which is important for VF termination. Here, we implemented the concept of post-shock synchronized pacing by a real-time feedback mechanism and demonstrated a successful VF termination by the post-shock SyncP strategy. Further optimization of this technique may prove effective in improving the defibrillation efficacy for low-energy ventricular defibrillation.