Riistama J, Lekkala J
Inst. of Meas. & Inf. Technol., Tampere Univ. of Technol., Tampere, Finland.
Conf Proc IEEE Eng Med Biol Soc. 2006;2006:6021-4. doi: 10.1109/IEMBS.2006.259712.
A research was made on electrical properties of implantable electrodes. Materials used for this research were platinum (Pt), stainless steel (SS), gold (Au) and silver-silver chloride (Ag/AgCl) that was used as a reference electrode material although it is not suitable for implantation, Wetzig et al. The interface impedance of the electrode-electrolyte-interface in three different electrolytes, saline (0.9 m-% NaCl), SBF (simulated body fluid) and PBS (phosphate buffered saline), was measured. A Randies equivalent circuit was used to model the electrode-electrolyte-interface. The two independent components of the electric equivalent circuit of the interface were quantified for all materials in all electrolytes. Impedance of the electrode-electrolyte interface was measured by a slightly modified technique presented originally by Lario-Garcia et al.
对可植入电极的电学特性进行了一项研究。本研究使用的材料有铂(Pt)、不锈钢(SS)、金(Au)以及用作参比电极材料的银 - 氯化银(Ag/AgCl),尽管它不适合植入(韦齐格等人)。测量了在三种不同电解质,即生理盐水(0.9 m-% NaCl)、模拟体液(SBF)和磷酸盐缓冲盐水(PBS)中电极 - 电解质界面的界面阻抗。使用兰迪等效电路对电极 - 电解质界面进行建模。对所有电解质中的所有材料,量化了界面等效电路的两个独立组件。电极 - 电解质界面的阻抗通过最初由拉里奥 - 加西亚等人提出的一种略有修改的技术进行测量。