Evaluation of contralateral breast skin doses by thermoluminescent dosimeters of patients receiving adjuvant radiotherapy for breast cancer.

作者信息

Gorken I B, Kentli S, Alanyali H, Karagüler Z, Kinay M

机构信息

Department of Radiation Oncology, Dokuz Eylül University Hospital, Izmir, Turkey.

出版信息

J BUON. 2002 Oct-Dec;7(4):351-4.

Abstract

PURPOSE

It is reported that low dose radiation received by the contralateral breast (CLB) during adjuvant radiotherapy (RT) is carcinogenic. This trial was planned to evaluate the CLB skin doses received during adjuvant RT of breast carcinoma.

PATIENTS AND METHODS

Twenty-four breast carcinoma patients treated locally or locoregionally with adjuvant RT were included. RT was performed with only tangential fields (TA) in 6 patients whereas 9 patients had an extra internal mammary (IM) field (TAIM). The remaining 9 patients received 5-field locoregional RT (5FLR). All patients were treated with wedge filters except for 3 TA patients. Of 9 5FLR patients IM fields were treated with Co60 in 5 and with electrons in the remaining 4 patients. LiF(2)-based Ribbon type thermoluminescent dosimeters (TLD) were used for dose evaluation. An average of 10 TLD's, placed with 1 cm gaps beginning from the medial border of the treatment field along the central axis were used to obtain dose measurements. Median measure of TLD's between 2-8 cm and maximum dose point (MDP) values in the same range were used to evaluate the CLB dose.

RESULTS

In TA patients the CLB skin received 6.3% of the total dose in patients treated with wedge filters and 7.13% with half-beam blocks. For 6 TAIM patients with IM fields treated with Co60, the CLB dose was 7.24%. In 5 of 9 5FLR patients, whose IM fields were treated with Co60 the CLB skin received 8.8% of the total dose, while for electron beam therapy the CLB dose was 5.44%. CLB median MDP values were as follows: 12.76% in TA patients treated with wedge filters and 11.45% with half-beam blocking; 11.89% in TAIM patients with IM fields treated with Co60 and 7.83% with electron beams; 12.29% in 5FLR patients of whose IM fields were treated with Co60 and 8.94% with electron beams.

CONCLUSION

When compared to wedge filters, halfbeam blocks caused 13% increase in CLB doses. If IM fields were added, 27.5% and 62% increases at CLB doses were established with Co60 when compared to electron beam RT in 3-field and 5-field treatments, respectively. CLB doses increased by 15-40% with the increased number of treatment fields. MDP values were also found to be higher with IM fields treated with Co60, but the number of treatment fields and accessories used seemed to have no effect on MDP doses. We conclude that by using wedge filters instead of half-beam blocks and by increasing the number of fractions treated with electron energies for IM fields, apparent decreases in CLB doses can be obtained. Large number of cases is needed to statistically establish the significant differences between subgroups.