Calderini Angelo, Maiorana Carlo, Garlini Giuliano, Abbondanza Tommaso
Department of Oral Surgery, School of Dentistry, University of Milan, Italy.
Int J Oral Maxillofac Implants. 2007 Sep-Oct;22(5):831-8.
To present a simplified method for assessment of precision of fit between screw-retained frameworks and supporting implants in clinical situations.
Torque-angle signature analysis is a method for analysis of tension within a joint. The OsseoCare device developed by Nobel Biocare was utilized for the tests. Three titanium frameworks were fabricated on the same master cast according to 3 different methods: (1) 1-piece casting, (2) the CNC (computer numeric controlled) method, and (3) the CTiP (Cresco Ti Precision) method. These frameworks were used to test the device and indirectly the application of the torque-angle signature analysis technology.
The frameworks fabricated according to the CNC and CTiP methods demonstrated OsseoCare tracings similar to the reference models of passively fitting joints, while the 1-piece cast framework did not.
According to this pilot study, the OsseoCare device and torque-angle signature analysis proved to be feasible for clinical assessment of fit between frameworks and supporting implants.