Portable TL dosemeter--ESD phantom combination for chest and lumbar spine radiography.

作者信息

Niittymäki Henri, Hakanen Arvi, Rautio Simo, Järvinen Hannu

机构信息

Department of Physics, University of Helsinki, Helsinki, Finland.

出版信息

Radiat Prot Dosimetry. 2008;130(2):224-7. doi: 10.1093/rpd/ncm492. Epub 2008 Jan 19.

Abstract

A thermoluminescence dosemeter (TLD)-entrance surface dose (ESD) phantom combination was calibrated in terms of air-kerma in IEC RQR X-ray radiation qualities between 50 and 150 kV. The ESD phantom was designed and constructed as a part of the work. With the combination, air-kermas were measured for four radiological examinations (two chest and two lumbar spine examinations in two hospitals), and converted to ESDa using Monte-Carlo calculated BSF data tabulated for different tube voltages, filtrations and beam diameters at 1 m distance. The results agreed with the ionisation-chamber measurements within the reported overall uncertainty of the TLD method. In the calibration, the ESD phantom can be replaced by the ISO water slab phantom unaffecting the reported overall uncertainty. Backscatter-related parameters for the ISO water slab phantom and the newly designed ESD phantom were determined for the IEC RQR qualities used in the secondary standard dosimetry laboratory of STUK at 1 m distance, including an approximate Hp(10)/Hp(0.07) ratio.

摘要

在50至150 kV的IEC RQR X射线辐射质条件下,对热释光剂量计(TLD)-体表剂量(ESD)体模组合进行了空气比释动能校准。ESD体模是作为这项工作的一部分设计和制作的。利用该组合,对四项放射学检查(两家医院的两次胸部检查和两次腰椎检查)的空气比释动能进行了测量,并使用针对不同管电压、滤过和1 m距离处射束直径编制的蒙特卡洛计算反向散射因子(BSF)数据,将其转换为体表剂量(ESDa)。结果在TLD方法报告的总不确定度范围内与电离室测量结果一致。在校准过程中,ESD体模可用ISO水模替代,而不会影响报告的总不确定度。在芬兰国家辐射与核安全管理局(STUK)二级标准剂量学实验室使用的1 m距离处,针对IEC RQR辐射质确定了ISO水模和新设计的ESD体模的反向散射相关参数,包括近似的Hp(10)/Hp(0.07)比值。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验