体层成像 PENN-PET 扫描仪的计数率性能。

The countrate performance of the volume imaging PENN-PET scanner.

机构信息

Dept. of Radiol., Pennsylvania Univ., Philadelphia, PA.

出版信息

IEEE Trans Med Imaging. 1994;13(4):610-8. doi: 10.1109/42.363107.

Abstract

The UGM PENN-PET camera uses large position sensitive detectors and operates without septa. This design results in high sensitivity and 3-D imaging capability, but poses problems in high countrate situations. The maximum true countrates and random countrates have been measured, as a function of object size in the field-of-view. The countrate performance is understood in terms of the limiting process rates and event rejection in the camera. In addition, the camera is calibrated to generate absolute activity concentrations to within 5% by correcting for system deadtime with up to 3 mCi in the field of view in brain studies (50% deadtime at 3 muCi/mL). This allows the performance of a variety of brain and body studies, and accurate quantitation of the data over a wide range of imaging countrates, from (18)F-FDG brain studies to (15)O-water dynamic brain studies.

摘要

UGM PENN-PET 相机使用大型位置灵敏探测器,且无隔室操作。这种设计可实现高灵敏度和 3D 成像能力,但在高计数率情况下会带来问题。已针对视场中的目标尺寸,测量了最大真实计数率和随机计数率。根据相机中的限制过程速率和事件拒绝情况,可了解计数率性能。此外,相机经过校正,可通过校正系统死时间,将视野中高达 3mCi 的放射性活度浓度绝对校正至 5%以内(在 3μCi/mL 时的死时间为 50%)。这使得各种脑和体研究得以实现,并可在从(18)F-FDG 脑研究到(15)O-水动态脑研究的广泛成像计数率范围内,对数据进行准确定量。