Maess Johannes, Fleming Andrew J, Allgöwer Frank
Institute for Systems Theory and Automatic Control, University of Stuttgart, Pfaffenwaldring 9, 70550 Stuttgart, Germany.
Rev Sci Instrum. 2008 Jan;79(1):015105. doi: 10.1063/1.2826428.
Piezoelectric tube scanners are widely used in scanning probe microscopes to position the sample or the probe. Fast and accurate scanning requires the suppression of dominant low-frequency resonances as well as the compensation of dynamics-coupling effects. The present article gives a detailed description of the fully coupled tube scanner dynamics over a wide frequency range modeled by finite element (FE) analysis using the commercially available software package ANSYS. The effect of a sample mass attached to the top of the tube is investigated by considering its added mass and local stiffening. A model order reduction scheme is applied to obtain a low order model that describes the lateral and vertical deflections as well as the voltage induced on quadrant electrodes. Comparison to experimental data demonstrates a good agreement for both the full FE model and reduced order model.