Kwon Min Jung, Lee Soo Youn
Department of Laboratory Medicine and Genetics, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Korea.
Korean J Lab Med. 2008 Feb;28(1):8-15. doi: 10.3343/kjlm.2008.28.1.8.
Point-of-care testing (POCT) glucometers are increasingly being used for making important therapeutic decisions and managing diabetes. We examined the analytical performance of GLUCOCARD X-METER (ARKRAY Global Business Inc., Japan) against three other glucometers and a reference laboratory method.
We evaluated the analytical performance of GLUCOCARD X-METER in comparison with three other glucometers. Studies on precision, linearity, the analysis time, and effects of hematocrit and temperature were carried out and the results were compared with those of the laboratory reference method (hexokinase method by Hitachi 760, Hitachi Co., Japan).
GLUCOCARD X-METER showed a good linearity and within-run and total-run precision. Comparison between each glucometer and the Hitachi 7600 showed a good correlation. Although differences with the reference method were within an allowable range, all glucometers showed variable bias. Application of an insufficient amount of blood could produce some changes in test results. Changes in hematocrit were found to cause overestimation or underestimation of glucose values. For some test strips, the results were affected by prolonged exposure to room temperature or 4 degrees C refrigerator.
GLUCOCARD X-METER showed a good analytical performance in linearity, precision, and comparison. The effect of hematocrit, sample volume, and storage condition for test strips were noted and glucometers had variable deviations to both directions from laboratory reference values (<20%). The GLUCOCARD X-METER provided rapid and reliable measurements of blood glucose. It could be appropriate for monitoring blood glucose values in diabetic patients.
即时检验(POCT)血糖仪越来越多地用于做出重要的治疗决策和管理糖尿病。我们将GLUCOCARD X-METER血糖仪(日本ARKRAY环球商业公司)与其他三款血糖仪以及一种参考实验室方法进行了分析性能比较。
我们评估了GLUCOCARD X-METER血糖仪与其他三款血糖仪的分析性能。进行了精密度、线性、分析时间以及血细胞比容和温度影响的研究,并将结果与实验室参考方法(日本日立公司日立760型的己糖激酶法)的结果进行了比较。
GLUCOCARD X-METER血糖仪显示出良好的线性以及批内和总批精密度。各血糖仪与日立7600之间的比较显示出良好的相关性。尽管与参考方法的差异在允许范围内,但所有血糖仪均显示出不同程度的偏差。采血量不足会使检测结果产生一些变化。发现血细胞比容的变化会导致葡萄糖值的高估或低估。对于一些试纸条,结果会受到室温或4℃冰箱长时间放置的影响。
GLUCOCARD X-METER血糖仪在线性、精密度和比较方面显示出良好的分析性能。注意到了血细胞比容、样本量和试纸条储存条件的影响,血糖仪与实验室参考值在两个方向上均存在不同程度的偏差(<20%)。GLUCOCARD X-METER血糖仪能够快速、可靠地测量血糖。它适合用于监测糖尿病患者的血糖值。