Standardization of magnetocardiography in nonhuman primates.

作者信息

Seki Yusuke, Muneyuki Kenta, Kandori Akihiko, Tsukada Keiji, Terao Keiji, Ageyama Naohide

机构信息

Advanced Research Laboratory, Hitachi Ltd. 1-280 Higashi-Koigakubo, Kokubunji, Tokyo 185-8601, Japan.

出版信息

Phys Med Biol. 2008 Mar 21;53(6):1609-18. doi: 10.1088/0031-9155/53/6/007. Epub 2008 Feb 25.

Abstract

To establish the electrophysiological mappings of nonhuman primates by using magnetocardiogram (MCG) data and obtain the normal values of MCG parameters, we used 64-channel superconducting quantum interference devices to measure 8x8 MCG data for 95 cynomolgus monkeys (Macaca fascicularis, 51 female and 44 male). The PQ interval, QRS duration, QT interval and QTc were respectively 79+/-14 ms, 42+/-7 ms, 222+/-23 ms and 363+/-25 (mean+/-SD), and these parameters did not differ significantly between female and male monkeys. These results indicate the normal values of the MCG parameters of the cynomolgus monkey and should facilitate animal experiments in magnetocardiography.