Zhang Wei, Wilson Christopher R, Danielson Neil D
Department of Chemistry and Biochemistry, Miami University, Oxford, OH 45056, USA.
Talanta. 2008 Feb 15;74(5):1400-7. doi: 10.1016/j.talanta.2007.09.016. Epub 2007 Sep 29.
An indirect fluorescence (FL) detection method via the reactivity of UV-photolyzed 2-phenylbenzimidazole-5-sulfonate (PBSA) has been developed for non-fluorescent aromatic compounds. At high pH with UV photolysis, PBSA in the excited state is known to be quenched by reaction with oxygen species and analyte compounds that are reactive toward these oxygen species produced during photolysis can lessen the loss of PBSA FL. After off-line photolysis of PBSA in the presence of various nitro-aromatic test compounds, the increase in PBSA FL is clearly evident. A flow injection (FI) instrument using a PBSA mobile phase propelled through a Teflon coil wrapped around a Hg lamp is optimized and modified for use for liquid chromatography (LC). For the on-line FI determination of the non-fluorescent nitro-aromatic compounds such as 4-nitroaniline, 2-nitrophenol, 3-nitrophenol, 4-nitrophenol, and alpha-nitronaphthalene, a positive linear response for PBSA FL from about 0.5 to 15 microM and detection limits generally between 0.2 and 1 microM (4-20 pmol) are found. Linear responses and detection limits of selected pharmaceutical compounds such as the antibacterial nitrofurantoin, antihistamines chlorpheniramine and brompheniramine, and other compounds were similar. In general, detection limits using UV detection at about 214 nm were not as good in the 1-2 microM range but linearity extended up to 100 microM. The amino acid phenylalanine and small peptides containing this aromatic amino acid were also determined using this method. Application of this detection method for the liquid chromatography determination of 4-nitroaniline, 2-nitrophenol, nitrofurantoin, and salicylate is shown.
通过紫外光解的2-苯基苯并咪唑-5-磺酸盐(PBSA)的反应性,开发了一种用于非荧光芳香族化合物的间接荧光(FL)检测方法。在高pH值和紫外光解条件下,已知处于激发态的PBSA会与氧物种反应而猝灭,而对光解过程中产生的这些氧物种具有反应性的分析物化合物可以减少PBSA荧光的损失。在各种硝基芳香族测试化合物存在下对PBSA进行离线光解后,PBSA荧光的增加明显可见。对一种使用PBSA流动相通过缠绕在汞灯周围的聚四氟乙烯线圈推进的流动注射(FI)仪器进行了优化和改进,以用于液相色谱(LC)。对于非荧光硝基芳香族化合物如4-硝基苯胺、2-硝基苯酚、3-硝基苯酚、4-硝基苯酚和α-硝基萘的在线FI测定,发现PBSA荧光在约0.5至15 microM范围内呈正线性响应,检测限一般在0.2至1 microM(4 - 20 pmol)之间。所选药物化合物如抗菌剂呋喃妥因、抗组胺药氯苯那敏和溴苯那敏以及其他化合物的线性响应和检测限相似。一般来说,在约214 nm处使用紫外检测时,1 - 2 microM范围内的检测限不太好,但线性范围可扩展至100 microM。还使用该方法测定了氨基酸苯丙氨酸和含有这种芳香族氨基酸的小肽。展示了该检测方法在液相色谱测定4-硝基苯胺、2-硝基苯酚、呋喃妥因和水杨酸盐中的应用。