Bisdas S, Bohning D E, Besenski N, Nicholas J S, Rumboldt Z
Department of Radiology, Medical University of South Carolina, Charleston, SC 29425, USA.
AJNR Am J Neuroradiol. 2008 Jun;29(6):1128-33. doi: 10.3174/ajnr.A1044. Epub 2008 Mar 27.
There is no reproducibility study of fractional anisotropy (FA) measurements at 3T using regions of interest (ROIs). Our purpose was to establish the extent and statistical significance of the interrater variability, the variability observed with 2 different b-values, and in 2 separate scanning sessions.
Twelve healthy volunteers underwent MR imaging twice. MR imaging was performed on a 3T unit, and FA maps were analyzed independently by 2 observers using ROIs positioned in the corpus callosum, internal capsules, corticospinal tracts, and right thalamus. Changes in FA values (x10(3)) measured with 2 b-values (700 and 1000 s/mm(2)), age-related differences, interobserver agreement, and measurement reproducibility were assessed.
In the right internal capsule genu (FA = 702/728; b = 1000/700 s/mm(2)) and the left anterior limb of the internal capsule (AIC; FA = 617/745; b = 1000/700 s/mm(2)), the FA values were significantly different between the 2 b-values (P = .02 and .05, respectively). Significant age-related differences in FA were observed in the genu of the corpus callosum and in the left AIC. Interrater measurements showed fair-to-moderate agreement for most anatomic structures. The lowest significant change for a single subject regarding any FA values between the 2 sessions was in the corpus callosum (4%), whereas the highest one was in the corticospinal tracts (27%). The Bland-Altman plot analysis showed that the 1000-s/mm(2) b-value gave satisfactorily reproducible measurements equally good or better than the 700-s/mm(2) b-value.
The reproducibility of FA estimates using ROIs was satisfactory. Measurements with a b-value at 1000 s/mm(2) showed superior reproducibility in most anatomic locations.
目前尚无关于在3T场强下使用感兴趣区(ROI)测量分数各向异性(FA)的重复性研究。我们的目的是确定不同评估者之间的变异性程度及其统计学意义,观察使用2种不同b值以及在2次独立扫描过程中的变异性。
12名健康志愿者接受了两次磁共振成像(MR成像)。在3T设备上进行MR成像,两名观察者使用位于胼胝体、内囊、皮质脊髓束和右侧丘脑的ROI独立分析FA图。评估了使用2种b值(700和1000 s/mm²)测量的FA值变化、年龄相关差异、观察者间一致性以及测量重复性。
在右侧内囊膝部(FA = 702/728;b = 1000/700 s/mm²)和左侧内囊前肢(AIC;FA = 617/745;b = 1000/700 s/mm²),2种b值之间的FA值存在显著差异(分别为P = 0.02和0.05)。在胼胝体膝部和左侧AIC观察到与年龄相关的FA显著差异。对于大多数解剖结构,观察者间测量显示出中等程度的一致性。在两次扫描之间,单个受试者任何FA值的最低显著变化出现在胼胝体(4%),而最高变化出现在皮质脊髓束(27%)。Bland-Altman图分析表明,1000 s/mm²的b值给出的测量重复性令人满意,与700 s/mm²的b值相当或更好。
使用ROI估计FA的重复性令人满意。在大多数解剖位置,b值为1000 s/mm²的测量显示出更好的重复性。