Khan W S, Agarwal M, Malik A A, Cox A G, Denton J, Holt E M
Musculoskeletal Research Group, University of Manchester, Oxford Road, Manchester M13 9PT, UK.
J Bone Joint Surg Br. 2008 Apr;90(4):502-5. doi: 10.1302/0301-620X.90B4.20302.
Metallosis after shoulder replacement has not previously been described in the literature. We report a patient who developed extensive metallosis after implantation of an uncemented Nottingham shoulder replacement. He underwent a revision procedure. Examination of the retrieved prosthesis showed that the titanium porous coating was separating from the humeral stem and becoming embedded in the ultra-high-molecular-weight polyethylene glenoid component, resulting in abrasive wear of the humeral component. There was metallosis despite exchange of the modular humeral head. Both components had to be exchanged to resolve the problem.
肩置换术后金属沉着病此前在文献中尚未有描述。我们报告一例患者,其在植入非骨水泥型诺丁汉肩假体后发生广泛金属沉着病。他接受了翻修手术。对取出的假体进行检查发现,钛多孔涂层与肱骨干分离并嵌入超高分子量聚乙烯关节盂部件中,导致肱骨部件出现磨蚀性磨损。尽管更换了模块化肱骨头,仍存在金属沉着病。两个部件都必须更换才能解决问题。