Wang C-K Chris, Seidaliev Marat, Mandapaka Anant
Nuclear & Radiological Engineering Program, Georgia Institute of Technology, Neely Research Center, 900 Atlantic Drive, Atlanta, GA 30332-0425, USA.
Health Phys. 2008 May;94(5):440-8. doi: 10.1097/01.HP.0000300493.53385.5d.
A novel tissue-equivalent proportional counter (TEPC) based on a gas electron multiplier (GEM) for measuring H*(10) for neutrons was designed and constructed. The pulse height spectra (PHS) of two different neutron sources (a 252Cf source and a AmBe source) were measured using the new TEPC. The measurements were made with the TEPC filled with two different gases (10P gas and a propane-based tissue-equivalent gas) at various pressures. A computer simulation of the new TEPC, based on the Monte Carlo method, was performed to obtain the PHS for the two neutron sources. It is shown that the experimental results agree well with the simulation results for both 252Cf and AmBe neutron sources. Several outstanding problems are discussed and suggestions are made to make the GEM-based TEPC a practical neutron rem meter. The potential advantage of this novel neutron rem meter would be its low weight and compactness.
设计并制造了一种基于气体电子倍增器(GEM)的新型组织等效正比计数器(TEPC),用于测量中子的H*(10)。使用新型TEPC测量了两种不同中子源(252Cf源和AmBe源)的脉冲高度谱(PHS)。在不同压力下,用填充两种不同气体(10P气体和丙烷基组织等效气体)的TEPC进行测量。基于蒙特卡罗方法对新型TEPC进行了计算机模拟,以获得两种中子源的PHS。结果表明,对于252Cf和AmBe中子源,实验结果与模拟结果吻合良好。讨论了几个突出问题,并提出了使基于GEM的TEPC成为实用中子剂量计的建议。这种新型中子剂量计的潜在优势在于其重量轻和体积小。