Huang Hui-Hsun, Lee Yi-Hui, Chan Hsiao-Lung, Wang Yong-Ping, Huang Chi-Hsiang, Fan Shou-Zen
Department of Anesthesiology, National Taiwan University Hospital and National Taiwan University College of Medicine, Taipei, Taiwan.
Med Biol Eng Comput. 2008 Oct;46(10):977-84. doi: 10.1007/s11517-008-0342-y. Epub 2008 Apr 15.
The measurement of anesthetic depth is important in anesthesiology. Although heart rate variability (HRV) is profoundly affected by general anesthesia, it has not yet been commonly used in this field. One of the reasons is the lack of suitable parameters of HRV for short-term observations. In this study, we designed a time domain parameter of HRV named the similarity index. It was based on observing the trend of the distribution of instantaneous heart rates as time moved. Taking epochs of ECG data as short as 64 s can derive the index. We observed the values of this index of 30 patients when they were awake and under isoflurane anesthesia. The values had very little overlapping between the two states and the prediction probability to distinguish the two states was 0.91. We suggest that HRV, if suitably treated, can play more roles in the monitoring of anesthetic depth.
麻醉深度的测量在麻醉学中很重要。尽管心率变异性(HRV)受到全身麻醉的深刻影响,但它尚未在该领域得到广泛应用。原因之一是缺乏适用于短期观察的HRV参数。在本研究中,我们设计了一种HRV的时域参数,称为相似性指数。它基于观察瞬时心率随时间变化的分布趋势。取短至64秒的心电图数据片段即可得出该指数。我们观察了30例患者在清醒和异氟烷麻醉状态下该指数的值。两种状态下的值几乎没有重叠,区分两种状态的预测概率为0.91。我们认为,如果对HRV进行适当处理,它可以在麻醉深度监测中发挥更多作用。